Hygosonic

Installation and Operating Instructions

9000-616-14/30

*9000-616-14/30*

Содержание Hygosonic

Страница 1: ...EN 2016 12SE Hygosonic Installation and Operating Instructions 9000 616 14 30 9000 616 14 30 ...

Страница 2: ......

Страница 3: ...pecial accessories 7 4 2 Disposable materials 7 5 Technical Data 7 6 Functional Description 8 Installation 7 Set up 9 7 1 Room for set up 9 7 2 Set up alternatives 9 8 Electrical Connections 9 9 Circuit diagram 9 Operation 10 Instructions for use 10 10 1 Performing routine checks of the unit before initial commissioning 11 10 2 Preparation 12 10 3 Operation pad 13 10 4 Cleaning process 14 10 5 Emp...

Страница 4: ...ed in this document are registered names The translation of these Installation and Operating Instructions has been carried out in good faith No liability can be accepted for mi stakes in the translation The enclosed German version of these Installation and Operating Instructions is the reference versi on If there is any doubt about the translation please consult your dealer Any reprinting of the t...

Страница 5: ...gulations must be observed Converting or modifying the appliance in any way is strictly prohibited In such cases any and all guarantees immediately become inva lid The operation of modified appliances can be punishable by law In the interests of trou ble free operation the operator is responsible for observing these regulations Before every use the operator must check the functional safety and the...

Страница 6: ...ion Do not use with flammable cleaning agents Danger of burns or scalding Danger housing bath and fluid can be come very hot Maintain max fluid levels Ultrasonic operation with enhanced performance Optimum ultra sound distribution in the fluid Degassing of fluid Ultrasonic operation On Off Temperature in C Time in minutes Wear protective gloves Wear protective glasses 3 1 Model identification plat...

Страница 7: ...te 6035 111 00 Instrument holding tongs 6030 012 00 4 2 Disposable materials Instrument disinfection ID 212 CDI212C6150 ID 212 forte CDI212F6150 ID 213 CDI213C6150 ID 220 Bath for drill heads CDI220C6150 MD 530 Cement remover denture cleaner CCA530C6150 MD 535 Gypsum remover CCA535C6150 5 Technical Data Hygosonic Voltage V 220 240 Frequency Hz 50 60 Ultrasonic frequency kHz 37 Power consumption es...

Страница 8: ...wave distribution without any formation of ultra sound shadows behind or beneath instruments which might be bundled together thus enhancing the cleaning effect To achieve the optimum cleaning results the Sweep function should be switched on Degas Auto Degas function 4 Fluids contain gaseous elements e g oxygen This function removes any interfering gas bub bles from the fluid Auto Degas When fresh ...

Страница 9: ...called wet rooms is not permitted 7 2 Set up alternatives Place the Hygosonic on a horizontal flat and dry working surface at an appropriate height pre ferably close to a correctly installed and earthed power outlet socket 8 Electrical Connections Before first operation check the supply voltage with the voltage requirements of the model identification plate Power Supply 220 240 V using an earthed ...

Страница 10: ...ntact with the floor of the bath itself therefore always ensure that cleaning is always carried out using the insert tray submersible tray or the special beakers Use the special beakers when using agents containing strong acids e g MD 530 MD 535 or alkali e g ID 220 Do not use any flammable fluids e g petrol solvents Do not use chemicals which contain or produce chloride ions this includes some di...

Страница 11: ...is perfo rated via cavitation during operation The perforations in the foil are photographed or scanned and compared with previous results The perforation and its distribution can be used to determine the efficiency The inspection must always be perfor med under identical conditions The in fluence factors to be determined are The filling height of the ultrasonic bath Time and intensity of the ultr...

Страница 12: ...ng using acidic or strong alkali solutions should only be carried out in the special glass beakers 12 Fill the solvents e g MD 530 into the beaker s Place the parts into the beaker 12 Place the punched steel cover 11 for the be akers into place Place the glass beakers complete with lids into the Hygosonic By increasing the temperature of the bath max 50 C the cleaning effect can be enhanced Procee...

Страница 13: ...anually 1 Set time on the rotating switch 70 to 0 2 Press key 5 Switch on heating Set temperature on the rotating switch 8 Switch off heating Set temperature on the rotating switch 80 to 0 Switch on Degas function 1 Set time on the rotating switch 7 2 Press key 5 3 Press Degas key 4 Switch off Degas function Press Degas key 4 ultrasound continues normally Switch on Auto Degas function Press Degas ...

Страница 14: ...0 Select times When using disinfectants be sure to observe the reaction times according to the manufacturer s instructions Switch on ultrasound with key 5 During operation never touch the cleaning fluids or any ultrasound parts If necessary switch on function Boost 2 or Sweep 3 After the cleaning process has finished place the tray 10 and glass beakers 12 onto the inner side of the cover 14 as sho...

Страница 15: ...solutions For order numbers see 4 2 Disposable materials For description see enclosed leaflet order number P007 080 01 D P007 080 02 GB 11 Maintenance 11 1 Routine check of the unit Perform the foil test at regular intervals The exact instructions for the foil test routine inspection of the Hygosonic are available in our download centre here www duerrdental com 11 2 Ultrasound bath care In order t...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ...DÜRR DENTAL SE Höpfigheimer Str 17 74321 Bietigheim Bissingen Germany Fon 49 7142 705 0 www duerrdental com info duerrdental com ...