Assembly
6
Requirements
The unit must not be set up or operated
within the vicinity of the patients (within a
radius of 1.5 m).
The unit can be installed either at the same level
as the surgery room or on a floor below (e.g. cel-
lar).
Due of the amount of noise generated, we rec-
ommend that the unit is installed in an adjoining
room.
The pipes provided on-site must at least meet
the country-specific requirements for drinking
water.
Further information can be found in our
separate planning information leaflet for
compressed air.
6.1
Installation/setup room
The room chosen for set up must fulfil the follow-
ing requirements:
– Closed, dry, well-ventilated room
– Should not be a room made for another pur-
pose (e. g. boiler room or wet cell)
– If the unit is installed in a machine room, e.g. in
an adjoining room or cellar, the requirements
set out in ISO
‑
TS 22595 must be complied
with.
ACHTUNG
Risk of overheating due to insufficient
ventilation
The units generates heat. Possibility of
heat damage and/or reduced service life
of the unit.
❯
Do not cover the unit.
❯
Install a fan for auxiliary ventilation in
rooms where ambient temperatures
exceed
³
40 °C while the unit is in
operation.
≥ 40 °C
6.2
Setup
The following conditions must be taken into
account for installation:
The air is filtered when it is sucked in. This
does not alter the composition of the air.
For this reason it is important to keep the
sucked-in air free of harmful substances
(e.g. do not suck in exhaust gases or
contaminated exhaust air).
– Clean, level and sufficiently stable subsurface
(note the weight of the unit).
– Type plate easy to read.
– Unit easy to access for operation and mainte-
nance.
– Easy-to-access power outlet to which the unit
is connected.
– Maintain sufficient distance to the wall (at least
20 cm).
– The compressed air pipe should be routed as
closely as possible to the place of installation
(note the length of the hose supplied).
6.3
Information about electrical
connections
❯
Ensure that the electrical connections to the
mains power supply are established in accor-
dance with current valid national and local reg-
ulations and standards governing the installa-
tion of low voltage units in medical facilities.
❯
Observe the current consumption of the
devices that are to be connected.
Assembly
9000-610-72/30 1907V007
15
EN