Connect the pressure hoses from the con-
trollers.
Additionally pull the hoses through the hose
from the accessories and secure the latter to
the pressure tank with a cable tie.
1
2
1
Auxiliary control pressure hose
2
Main control pressure hose
8.3
Network connection
Purpose of the network connection
The network connection is used to exchange
information or control signals between the unit
and a software installed on a computer, in order
to, e. g.:
– Display parameters
– Select operating modes
– Indicate messages and error situations
– Change unit settings
– Activate test functions
– Transmit data for archiving
– Provide documents concerning the units
During initial installation, we recommend a
router or server with DHCP so that the
unit is detected in the network.
❯
Plug the network cable into the control and into
a network socket.
❯
Connect to the computer network with the net-
work cable.
DHCP
compressor
8.4
Electrical connections
Safety when making electrical connections
The unit has no main power switch. For
this reason it is important that the unit is
be set up in such a way that the plug can
be easily accessed and unplugged if
required.
❯
The unit must only be connected to a correctly
installed power outlet.
❯
Make sure that none of the electrical cables
leading to the unit are under any mechanical
tension.
❯
Before initial start-up check that the mains sup-
ply voltage and the voltage stated on the type
plate match (see also "4. Technical data").