13.3 Preparation at the operating
location
Wear protective gloves.
Wear protective goggles.
Use a mask.
Use protective clothing.
WARNING
Risk of infection from contaminated
products
Danger of cross contamination
❯
Reprocess the product correctly and
promptly before its first use and after
every subsequent use.
Rinsing the handpiece with water
❯
Fill the fluid container to approx. 1/3 with water.
❯
Place the handpiece in a sink or in a suitable
container.
❯
Start the rinsing process:
Press the RINSE button for at least 2 seconds.
RINSE
❯
The LED flashes – the rinsing process takes
around 30 seconds and ends automatically.
Pre-cleaning
Do pre-cleaning of handpiece and acces-
sories no more than 15 minutes after the
unit has been used.
❯
Clean the exterior surfaces completely with two
cleaning cloths
. Make sure that the surfa-
ces are sufficiently moistened.
❯
Note the action time of the cleaning agent.
❯
Perform the procedure twice.
❯
Draw 3 x 20 ml cold water (temperature
< 20 °C) into a conventional sterile 20-ml dis-
posable pipette with Luer connection and rinse
the inner lumen of the handpiece.
20 ml
Transport
❯
Protect the device from contamination when
you transport it from the treatment chair to the
reprocessing location.
13.4 Dismantling the handpiece
❯
Unscrew the instrument, "Inserting/changing
instruments".
❯
Take off the removable parts of the handpiece,
"Disassembly".