Error messages / troubleshooting
Se
qu.
no.
Message
Cause;
(●) available actions
by whom
[001] System start (switched on)
The system has been switched on.
●
No action necessary
[002] System start (external reset)
A system start was triggered by the internal reset
logic.
●
No action necessary
[003] System start (watchdog reset) A system start was triggered by the internal watch
dog timer. A deliberate soft reset will also trigger
this message!
●
If this error occurs frequently, there is a fault in
the electronics, and the measuring head must
be sent in to the Service Dept for repair.
Manufacturer
(send the device
back)
[004] System start (BOD reset)
A system start has been triggered due to detection
of an undervoltage.
●
Check the 24V power supply
Specialised
electrician
[005] System start (with default
jumper)
The measuring head has been restarted by placing
the default jumper on to X14
●
No action necessary
[006] Zero point measurement is be
ing performed
The system is performing zero point measurement
●
No action necessary
[007] Reference point measurement
is being performed
The system is performing reference point measure
ment
●
No action necessary
[008] Contamination measurement
is being performed
The system is performing contamination measure
ment
●
No action necessary
[009] Parameter change saved
Changed parameter have been saved
●
No action necessary
[010] Logbook was reset
The records logbook has been cleared down
●
No action necessary
[011] Message counter was reset
The message counter has been reset
●
No action necessary
[012] Device starting
The device is in start mode and not yet ready to op
erate. The duration of the wait until readiness to op
erate is approx. 20 sec
●
No action necessary
[013] Linearity test being performed The system is in linearity measurement mode
➔ No action necessary
[014] LED comparison is being per
formed
The system is performing an LED comparison (initi
ated by a Service Engineer)
●
No action necessary
[015] Initial contamination measure
ment being performed
The system is performing an initial contamination
measurement (initiated by a Service Engineer)
●
No action necessary
8.3
Содержание D-R 290
Страница 12: ...12 D R 290 ...
Страница 18: ...1 General 18 D R 290 ...
Страница 20: ...20 D R 290 ...
Страница 28: ...2 Safety 28 D R 290 ...
Страница 30: ...30 D R 290 ...
Страница 38: ...3 Delivery 38 D R 290 ...
Страница 40: ...40 D R 290 ...
Страница 60: ...4 Product description 60 D R 290 ...
Страница 106: ...106 D R 290 ...
Страница 152: ...152 D R 290 ...
Страница 174: ...8 Maintenance 174 D R 290 ...
Страница 176: ...176 D R 290 ...
Страница 177: ...9 Appendix D R 290 177 Appendix Measurement point questionnaire Fig 9 1 Measurement point questionnaire 1 9 9 1 ...
Страница 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Страница 188: ...9 Appendix 188 D R 290 Fig 9 11 Dimensional diagram D TB 100 supply unit Dimensional diagram purge air unit D BL 9 3 7 ...
Страница 193: ...9 Appendix D R 290 193 Fig 9 18 Dimensional diagram weather protection cover control electronics protective device ...
Страница 194: ...9 Appendix 194 D R 290 ...
Страница 200: ...11 Index 200 D R 290 ...
Страница 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...