Analog Output [V]
Pressure
0.00
LO /
P
≤
0 mTorr
0.50
0.10 mTorr
1.00
1.0 mTorr
1.50
10.0 mTorr
2.00
100 mTorr
2.50
1.00 Torr
3.00
10.0 Torr
3.50
100 Torr
4.00
1000 Torr
4.52
HI / OFF
/
P
≥
1 x 10
4
Torr
Table 3
. Analog output with displayed pressure values
Pressure as a function of the analog output can be approximated by
P
= 0.01 * 10
2
V
where
V
is the analog output in volts and
P
is pressure in mTorr. For example, if
V
is equal to
3.50 V,
P
is approximately 100 Torr.
Pressure values are converted from torr to millibar or Pascal
units.
1 kΩ source impedance for analog output
Troubleshooting
Problem
Possible Cause
Diagnostic
Unit fails Self-Test
N/A
Restart unit; if restart fails,
contact Duniway Stockroom
Fuse(s) repeatedly
burn out
Incorrect AC input
voltage
Verify AC voltage; if unit fails,
contact Duniway Stockroom
Display is dim and reads
incorrect pressure values
Incorrect AC input
voltage
Verify AC voltage; if unit fails,
contact Duniway Stockroom
Incorrect VAC / ATM
values
Faulty pressure gauge
Verify pressure gauge is
operational
Incorrect VAC / ATM
values
Faulty power supply
Verify unit output voltage