![Drucker Diagnostics DASH COAG Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/drucker-diagnostics/dash-coag/dash-coag_operators-manual_2533145015.webp)
Drucker Diagnostics – Customer Service: +1-814-692-7661 –
G
ENERAL
S
PECIFICATIONS
The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
Tube Capacity
12 tubes,
up to 75 mm (4 mL)
Dimensions (W x D x H)
8 in x 11 in x 13 in
(20 cm x 28 cm x 34 cm)
Weight
12 lbs (5.4 kg)
Sound Level
61 dB A
Environmental Range
16 – 32°C
Voltage
95 – 253 VAC
Frequency
50/60 Hz
Power Requirement
220 Watts
Centrifuge Motor
½ H.P. Brushless DC
Max G-Force
4,400 xg
Max Speed
6,600 RPM
Min Cycle Time
1 minute
Max Cycle Time
30 minutes
Use only with approved accessories from the original manufacturer. A complete list of accessories is available
at druckerdiagnostics.com.
C
ALCULATING THE
G-F
ORCE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which can be calculated if you know
the RPM and the radius. Use the formula below or go to
www.druckerdiagnostics.com/g-force-calculator/
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)
2
In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)
2
Radius
3.5 in (9 cm)