Drucker Diagnostics DASH COAG Скачать руководство пользователя страница 13

Drucker Diagnostics – Customer Service: +1-814-692-7661 – 

[email protected]

 | Page 7 

T

ROUBLESHOOTING

 

NOTE: The latch must be turned completely clockwise to its stop  position for the centrifuge to operate. 

The centrifuge does not run

 

 

o

 

Verify that the centrifuge is powered. One of the LED lights 

should be on.  

o

 

Make sure the lid latch is turned completely clockwise to its stop 

position.  

o

 

If the centrifuge still does not run, contact Customer Service. 

The rotor does not spin freely 

o

 

Make sure nothing has fallen into the rotor chamber, 

following the procedure above. 

 

o

 

If nothing obstructs the rotor, the rotor may be damaged. 

Contact Customer Service for further assistance. 

The centrifuge makes a rattling 

noise when running 

o

 

Stop the centrifuge. Open the lid.  

o

 

Wearing PPE, remove tubes and tube holders/buckets and 

look for fallen objects or debris.  Carefully reach inside the 

rotor chamber with a tool to remove them.  

o

 

Inspect the rotor, tube holders or buckets for damage.  

o

 

If the tube holders or buckets have any damage, even slight, 

safely dispose of them and replace them.  

o

 

If the rotor appears damaged, contact Customer Service for 

further assistance. 

Excessive noise or vibration when 

the centrifuge is running

 

o

 

Verify that all four centrifuge feet are properly seated on a flat 

surface. 

o

 

Insure that the load is balanced according to instructions in the 

“Balancing Loads” section of this manual. 

o

 

Make sure that nothing has fallen into the rotor chamber.

 

The centrifuge stops and 

beeps continuously 

 

The load is not balanced. Press the UNLOCK button, open the lid, 

and balance the load as recommended elsewhere in this manual.

 

The centrifuge does not unlock after 

a run is completed

 

 

o

 

Wait until the rotor has come to a complete stop. If the lid knob 

still cannot be rotated, press the UNLOCK button and try again.  

o

 

If no LED light is on, the unit is 

not powered and the lid will not 

unlock by conventional means. 

Remove the latch label and use a 

pen to manually disengage the 

locking mechanism. Pull the 

mechanism towards the control 

panel and then unlatch and open 

the lid. 

 

o

 

If the unit is damaged, contact your authorized dealer or 

Drucker Diagnostics. 

 

Содержание DASH COAG

Страница 1: ...DASH COAG Operator s Manual...

Страница 2: ......

Страница 3: ...ial Setup 2 Quick Start 3 Settings 4 Review Cycle Time and Speed Settings 4 Changing Cycle Time and Speed Settings 4 Operation 5 Balancing Loads 5 Care and Preventative Maintenance 6 Cleaning and Disi...

Страница 4: ......

Страница 5: ......

Страница 6: ......

Страница 7: ...to reinforce the tube holders and provide high strength and durability o A clear lid permits safe observation of samples and optical calibration of speed o The lid safety system only allows entry into...

Страница 8: ...repairs have been performed This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with this operator manual may cause interference to radio c...

Страница 9: ...ng the locking mechanism Entry is only possible when the rotor is stopped Stop Pressing the UNLOCK button during operation will terminate the run and unlock the lid after the rotor has come to astop C...

Страница 10: ...in the current cycle 10 beeps flashes equal 10 minutes of run time Run time starts when the rotor reaches 90 of desired speed and stops when the rotor starts decelerating o Pressing the START button a...

Страница 11: ...will slow to a complete stop and the lid light will flash o The locking mechanism will disengage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed...

Страница 12: ...tingor storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is...

Страница 13: ...mage o If the tube holders or buckets have any damage even slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise...

Страница 14: ...o change the run preset time The lid does not open o Ensure that the lid knob is turned fully counterclockwise o If the knob cannot be turned counterclockwise turn it fully clockwise press UNLOCK and...

Страница 15: ...0 Watts Centrifuge Motor H P Brushless DC Max G Force 4 400 xg Max Speed 6 600 RPM Min Cycle Time 1 minute Max Cycle Time 30 minutes Use only with approved accessories from the original manufacturer A...

Страница 16: ...ectronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protect...

Отзывы: