Drucker Diagnostics 642B Скачать руководство пользователя страница 10

Drucker Diagn1-814-692-7761 - [email protected]

 

Page | 8

  

C

ARE 

A

ND 

P

REVENTATIVE 

M

AINTENANCE

 

With proper care and maintenance your centrifuge will provide years of 
laboratory service. For proper care, the following steps should be taken: 

1.

 

Provide Adequate Ventilation:

 For cooling purposes, the Model 

642B draws in ambient air through the air intake cover on the 
top of the lid and exhausts this air in the rear of the base. The 
centrifuge should be placed on a hard smooth surface for good 
air circulation. 

2.

 

Always Spin Balanced Loads:

 Make certain that you are always 

spinning a balanced load. The Model 642B has a unique counter 
balanced motor mounting design which, along with its rubber 
suction feet, produces excellent vibration dampening. However, 
out–of–balance loads may break glass test tubes and may 
produce unsatisfactory separation results. Proper load balancing 
will improve sample separation and extend the life of the 
centrifuge. Refer to page 7 on balanced loads for additional 
information on balancing the load. 

3.

 

Keep the Tube Holders Clean:

 NOTE: Always follow the safety 

guidelines of your laboratory to properly clean up and/or dispose 
of materials in the event that a substance known to be 
potentially toxic, radioactive or contaminated with a pathogenic 
microorganism is spilt in or on the centrifuge. Small glass 
fragments left in the tube holder after a tube breakage may 
adhere to the next test tube inserted in that holder. When this 
tube is handled, these fragments may puncture protective gloves 
and lacerate the operator’s fingers or hand. Remaining fragments 
may provide stress points on subsequent tubes and result in 
additional breakage. If a tube breakage occurs, carefully remove 
the tube holder. Properly dispose of the sample and tube 
fragments and thoroughly clean both the inside and outside of 
the tube holder. Insert a new tube cushion (if necessary) and 
replace the tube holder in the rotor. 

4.

 

Motor and Electrical Maintenance:

 The Model 642B uses a 

brushless permanent split capacitor AC motor. It should not need 
routine servicing for the life of the centrifuge. The electrical 
components are selected for high reliability and should not need 
routine service. 

5.

 

Tube Holder Replacement:

 It is recommended that the tube 

holders be replaced after 24 months of use. Inspect tube holders 
regularly for cracks. If cracks are discovered, replace 
immediately. 

Содержание 642B

Страница 1: ...642B Operator s Manual P N 03 0 0002 0078 Rev H...

Страница 2: ...Setup Location 4 Initial Setup Procedure 5 Operation 6 Spinning Balanced Loads 7 Rotor Removal And Installation 7 To remove the rotor 7 To install the rotor 7 Care And Preventative Maintenance 8 Clean...

Страница 3: ...o Seal accessory for the model is the non aerosol shield cap with appropriate tube holders contact Drucker Diagnostics to purchase In the event that materials of a higher risk group are being used mor...

Страница 4: ...through the transparent lid In the event that the lid is opened during a run the power to the motor is disconnected This general purpose laboratory centrifuge may also be used to spin approved contain...

Страница 5: ...at prevents the centrifuge from operating unless the lid is closed and latched Removable rotor for easy cleaning Brushless permanent split capacitor AC motor Clear lid for safe observation of samples...

Страница 6: ...he center No person or hazardous material shall be permitted in the clearance envelope during operation The operator time within the envelope shall be limited to the time necessary for loading unloadi...

Страница 7: ...ers inside the rotor and verify that they are seated properly 4 Close the lid Rotate the lid knob clockwise to its complete stop position 5 Turn the centrifuge on by turning the timer to 10 minutes 6...

Страница 8: ...ders Be sure to follow the rules for balanced loads page 7 4 Close the lid and turn the lid knob clockwise to its complete stop position 5 Turn on the machine by turning the timer to the desired run t...

Страница 9: ...nplug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning i...

Страница 10: ...safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microorg...

Страница 11: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Страница 12: ...is running o Check to see that the load is balanced o Make sure that nothing has fallen into the rotor chamber o Make sure that the nut in the center of the rotor is tight o Have a technician test the...

Страница 13: ...e to a complete stop and then unlock for 60 seconds If additional unlock time is needed press the OPEN STOP button with the machine plugged in and the rotor stopped If the lid remains locked after thi...

Страница 14: ...the top of the cabinet by a latching knob and pawl system When the knob is rotated clockwise the pawl grips the underside of the cabinet opening and prevents the lid from opening A mechanical stop pos...

Страница 15: ...ce horizontal 02 005 1 0006 Motor Assembly 1 30 H P 115 V A C permanent split capacitor 7729009 Capacitor 5uF 250V A C 7751043 Circuit Breaker 4A 7724071 Hinge friction 7732018 Seal lid gasket 02 002...

Страница 16: ...and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment For more...

Отзывы: