Operator’s Manual 

 

 

 

 

 

 

 

 

P/N 03-0-0002-0229 Rev. C 

Содержание 614VET

Страница 1: ...614VET Operator s Manual P N 03 0 0002 0229 Rev C...

Страница 2: ...5 Initial Setup Procedure 6 Control Panel 7 Verifying The Preset Time 8 Changing The Preset Time 8 Operation 9 Spinning Balanced Loads 10 Rotor Removal And Installation 10 To remove the rotor 10 To i...

Страница 3: ...e holders inside machine MODEL DESCRIPTION The Model 614VET is a continuous duty electronically controlled Fixed Angle centrifuge with a lid safety interlock system The unit is controlled by an electr...

Страница 4: ...flow design that prevents overheating of samples Heavy gauge steel construction for safety and durability Lid safety switch that prevents the centrifuge from operating unless the lid is closed and la...

Страница 5: ...uipment is present 2 Choose a setup location which meets the following criteria a A clearance height of 20 is required to open the lid b The clearance envelope is the space around the centrifuge which...

Страница 6: ...the rotor and verify that they are seated properly 6 Close the lid Rotate the lid knob clockwise to its complete top position The LATCHED indicator light should be illuminated If it is not make sure...

Страница 7: ...for fifteen 15 additional seconds After the centrifuge has passed this procedure it is ready for operation CONTROL PANEL RUNNING Lights up when the machine is in operation power is being applied to t...

Страница 8: ...CHANGING THE PRESET TIME NOTE Your centrifuge must be plugged in a Push the OPEN STOP button to disengage the lock and then open the lid b Push and hold the START and OPEN buttons for approximately th...

Страница 9: ...of two 2 minutes To display or change this time setting refer to page 8 6 Turn on the machine by pushing the START button on the control panel 7 The centrifuge should begin to spin The RUNNING indicat...

Страница 10: ...plug the centrifuge from the electrical outlet to eliminate the possibility of electrical shock or other injury 2 Remove the test tube holders 3 Remove the nut in the center of the rotor by turning it...

Страница 11: ...e safety guidelines of your laboratory to properly clean up and or dispose of materials in the event that a substance known to be potentially toxic radioactive or contaminated with a pathogenic microo...

Страница 12: ...in water or other cleaning solutions as this will cause damage and void the warranty 3 ONLY isopropyl alcohol soap and water or a 10 5500 PPM bleach solution should be used for cleaning and disinfect...

Страница 13: ...replace it if necessary The centrifuge does not run o Check the electrical outlet o Make sure the lid latch is turned completely clockwise to its stop position When the lid is closed properly the lat...

Страница 14: ...t to the desired length Check the run preset by following the instructions on page 8 If the preset is not the desired length follow the procedure on the same page to change the run preset time EMERGEN...

Страница 15: ...14VET has a true 0 RPM lid locking system The lid safety locking system keeps the lid locked at all times even during power failure and requires that the rotor be at rest in order to unlock the lid Th...

Страница 16: ...d 7751043 Circuit Breaker 7760002 Power cord 03 1 0008 0016 Pawl latch lid 03 1 0008 0009 Knob latch lid 7724071 Hinge friction 7732018 Seal lid gasket 7713032 125mm Tube Holder Black 02 002 1 0027 Li...