3.  Installation & Operation

3.1  General Requirements

When the sensor is received, check the completeness.
To identify the electrical and pressure connections, refer to the product datasheet or, if applicable, 

the specification drawing.
Do not use force when installing the sensor. Use only a wrench on the sensor’s hexagonal flats 

to tighten the sensor.
The ambient temperature and the process media to be measured must not exceed the ranges 

specified in the sensor specification.
Do not use the sensor where the pressure medium may freeze. This may cause damage to the 

sensor and connected pressure equipment.
The materials used for the primary enclosure and pressure bearing surfaces are identified in the 

product datasheet or, if applicable, the specification drawing. Make sure that the materials are 

applicable for the installation.
Before using the equipment, remove the plastic/rubber protection cap from the pressure 

connector.
Some models feature a white PTFE vent filter in the wall of the enclosure. Make sure the vent 

filter is correctly installed and is flush with the enclosure body.

3.2  Safety Measures

The operation of sensors in systems whose pressure may exceed the overload values specified 

in the data sheet or customer-specific specification drawing is not allowed.
Connection and detachment of sensors from the mains supplying the pressure of the medium to 

be measured must be done after the shutoff valve is closed from the process and the pressure in 

the working chamber is made equal to atmospheric.
The connecting pipes must have a one-way slope (not less than 1:10) from the pressure collection 

point up to the sensor, if the medium to be measured is gas, and down to the sensor if the medium 

is liquid. If this is not possible, when measuring gas pressure at the lower points of the connecting 

lines, it is necessary to install sludge vessels, and when measuring the liquid pressure at the 

highest points, install gas collectors.
Selected devices for mounting sensors should be mounted on straight sections, at the maximum 

possible distance from pumps, locking devices, elbows, expansion joints and other hydraulic 

devices. It is especially not recommended to install sensors in front of the shut-off device if the 

CAUTION  

Until installation, keep the unit in the original container with all the 

covers in position. The container and covers prevent contamination and 

damage. When not in use, keep the connections covered.

WARNING  

High pressures, temperatures and potentially poisonous pressure 

media are dangerous, and can cause injury to personnel and damage to property 

and the environment. Ensure correct installation, sealing of pressure interfaces 

and connection of the equipment. Ensure correct operation of the equipment in 

accordance with the specification. Use the applicable protection and obey all 

safety precautions.

WARNING  

For sensors intended for use in Explosive Atmospheres refer to 

additional instructions on Hazardous Area Installation.

Содержание UNIK 5 00 Series

Страница 1: ...Druck com UNIK 5 00 Series Pressure Sensing Platform Instruction Manual ...

Страница 2: ......

Страница 3: ...issued to protect the user and the equipment from injury or damage Use qualified personnel and good engineering practice for all procedures in this publication Maintenance The sensor must be maintained using the manufacturer s procedures and these should be carried out by authorized service agents or the manufacturer s service departments Druck com For technical advice contact the manufacturer WAR...

Страница 4: ...ired the extraction and use of natural resources for its production It may contain hazardous substances that could impact health and the environment In order to avoid the dissemination of those substances in our environment and to diminish the pressure on the natural resources we encourage you to use the appropriate take back systems Those systems will reuse or recycle most of the materials of you...

Страница 5: ...on 4 3 1 General Requirements 4 3 2 Safety Measures 4 3 3 Connecting to a Pressure Source 5 3 3 1 Media Compatibility 6 3 3 2 Pressure Containment 7 3 4 Power Requirements 8 3 5 Maintenance 9 3 5 1 Visual Inspection 9 3 5 2 Cleaning 9 3 6 Returned Goods Procedure 9 3 6 1 Safety Precautions 9 3 6 2 Important Notice 9 3 7 Electromagnetic Compatibility 9 3 7 1 Power Supply and Metering 9 3 7 2 Cable ...

Страница 6: ...Copyright 2022 Baker Hughes Company iv UNIK 5 00 Series English ...

Страница 7: ...tong International Industrial Park No 8 Xihu Road Wujin High Tech Industrial Zone Changzhou Jiangsu 213164 China Each sensor is marked to indicate the country of manufacture 2 Description 2 1 Purpose UNIK pressure sensors hereinafter referred to as sensors are designed for continuous measurement and conversion of gauge atmospheric absolute and differential pressure into an analogue output signal o...

Страница 8: ...ss to metal seal for electrical connections and a fluid filled cavity containing a silicon diaphragm with diffusion implanted resistors arranged in the form of a Wheatstone bridge The operating principle of the pressure sensor is based on the piezo resistive effect a change in resistance when applying pressure When pressure is applied the silicon diaphragm flexes changing the resistance of the imp...

Страница 9: ...ccuracy description Industrial Improved Premium 5 Model number 6 Serial number 7 Pressure range limits and unit of measurement 8 Input supply voltage range and current limits Note The symbol indicates the use of direct current 9 Output voltage or current range 10 Ambient temperature range 11 Reserved for applicable certification markings Position on sensor may vary 12 Manufacturer s name and addre...

Страница 10: ...ssure of the medium to be measured must be done after the shutoff valve is closed from the process and the pressure in the working chamber is made equal to atmospheric The connecting pipes must have a one way slope not less than 1 10 from the pressure collection point up to the sensor if the medium to be measured is gas and down to the sensor if the medium is liquid If this is not possible when me...

Страница 11: ...impact shock mechanical and thermal stresses Do not install the equipment where it can be damaged by a material that causes corrosion Provide additional protection for the equipment if it may be damaged in service When installing power supply and signal wiring the possibility of condensate entering the sensor cable entry should be avoided 3 3 Connecting to a Pressure Source When mounting the senso...

Страница 12: ...ical connector option 3 immersion fluid must be compatible with polyurethane For electrical connector option 4 immersion fluid must be compatible with Hytrel For electrical connector option N immersion fluid must be compatible with Kynar 5600 0 200 bar 0 2900 psi Fluids compatible with stainless steel 316L and Hastelloy C276 for differential versions only 201 500 bar 2915 7250 psi Liquids and grou...

Страница 13: ...rsion fluid must be compatible with nitrile butadiene rubber 5800 5900 0 200 bar 0 2900 psi Fluids compatible with stainless steel 316L and Hastelloy C276 201 500 bar 2915 7250 psi Liquids and group II gases compatible with stainless steel 316L 501 700 bar 7265 10150 psi Group II liquids and group II gases compatible with stainless steel 316L Table 2 Pressure Containment Product Pressure Range Pre...

Страница 14: ...DCR Passive 2 5 to 12 Vdc PDCR Linearized 7 to 12 Vdc 5600 5700 PTX 7 to 32 Vdc 5800 5900 PMP Basic 7 to 32 Vdc 12 to 32 Vdc for 0 to 10 V Output PMP Basic Configurable Maximum Output 1 V to 32 Vdc PTX 7 to 32 Vdc PDCR Passive 2 5 to 12 Vdc PDCR Linearized 7 to 12 Vdc Table 4 Supply Consumption Product Output Type Current Consumption 5000 PMP Basic 3 mA PMP Ratiometric 3 mA PMP Basic Configurable ...

Страница 15: ...n contact with hazardous or toxic materials Please supply the applicable Control of Substances Hazardous to Health COSHH or Material Safety Data Sheet MSDS references and precautions 3 6 2 Important Notice Service or calibration by unauthorized sources will affect the warranty and may not guarantee further performance If the equipment has hazardous area approval the approval will also be invalid 3...

Страница 16: ...t through which it is going to run Screened cable should always be used where electrical noise is present Good cabling practice will be reflected in signal quality 3 7 3 Earthing Grounding For the screening of the cable to be effective it is essential that the screen or drain conductor is permanently bonded to earth ground This should take place at the monitoring end of the cable as close to the p...

Страница 17: ......

Страница 18: ...s Company This material contains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are trademarks of their respective holders Office Locations Services and Support Locations ...

Отзывы: