Main Calibration

This selection uses the pressure and vacuum controllers of the ADTS to generate the 

required pressures for calibration. The necessary AIM values and RATE of pressure change 

can be set. A calibration standard of the required accuracy, connected to the output ports, 

measures the pitot and static channels.
This is the recommended method of calibration.

External:
This selection uses a deadweight tester (or similar calibration standard) to generate the 

pressure and vacuum for calibration. The ADTS, in measure mode, displays the applied 

pressure and vacuum in the pitot and static channels.

4.6.3  Select Calibration Mode

Select calibration mode from:

Static (Ps) channel - absolute
This mode calibrates the static (Ps) channel.

Connect the calibration standard to the static (Ps) output on the front panel. Fit the blank on the 

pitot (Pt) output.

Pitot (Pt) channel - absolute
This mode calibrates the pitot (Pt) channel, using an absolute pressure, calibration standard.

Connect the calibration standard to the pitot (Pt) output on the front panel.
Fit the blank on the Static (Ps) output.

Combined Static (Ps) and Pitot (Pt) channel - absolute
This mode enables combined calibration of static (Ps) and pitot (Pt) channels to minimize 

calibration time and errors due to calibration standard non-repeatability.

Connect the calibration standard to the pitot (Pt) output on the front panel.
Fit the blank on the Static (Ps) output.

When using combined static (Ps) and pitot (Pt) calibration, the pressures must be applied in 

ascending order starting from the lowest pressure. Each entered exact pressure will be used as 

a calibration point for both static (Ps) and pitot (Pt) until [END CAL] is selected. At that point the 

static (Ps) calibration will be completed, a valve will close to protect the static (Ps) channel and 

the remaining points will be used for the pitot (Pt) channel only.

Calibration Points

1. Following calibration initialization, the calibration entry screen appears, showing the present 

measured static (Ps) and pitot (Pt) values.

2. Two functions are common to INTERNAL and EXTERNAL pressure source:

[ACTUAL VALUE]

This allows the actual value of the pressure, applied by or measured by the calibration 

standard, to be entered. The pressure must be allowed to stabilize before entering the value. 

The minimum number of calibration points is two.

There is no upper limit to the number of points that may be entered. For specification 

accuracies to be achieved, Druck recommend a minimum of five points on static (Ps) with a 

further three on pitot (Pt) if both are calibrated together.

CAUTION  

Unless the [END CAL] instruction is given, the static PS sensor will 

be over-pressured with pitot PT values.

Содержание ADTS 403F

Страница 1: ...Druck com ADTS 403 405 R F Air Data Test System Calibration Manual...

Страница 2: ......

Страница 3: ...e condition The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage Use qualified personnel and good engineering practice for all pro...

Страница 4: ...an approved organization that collects and or recycles waste electrical and electronic equipment For more information contact one of these Our customer service department Druck com Your local governme...

Страница 5: ...al K0154 ADTS 405 IEEE 488 Option User Manual K0157 ADTS 405 SCPI User Manual Abbreviation Description A Ampere abs Absolute AC Alternating current ADTS Air data test system ALT Altitude ARINC Aeronau...

Страница 6: ...Pressure Total Pitot P W Password Qc Differential pressure SCPI Standard commands for programmable instruments TAS True airspeed TPM Test program manager Druck software package V Volts ve Positive ve...

Страница 7: ...ticular state or condition especially in a state of efficiency or validity Make sure To confirm that a proper condition exists to find out with certainty Operate Ensure that an item or system function...

Страница 8: ...Copyright 1996 Baker Hughes Company vi ADTS 403 405 Calibration Manual English...

Страница 9: ...Combined Pitot and Static Channels 11 4 6 Main Calibration 12 4 6 1 To Operate Calibration Switch 12 4 6 2 Pressure Source 12 4 6 3 Select Calibration Mode 13 4 7 Calibration and Adjustment of the Sta...

Страница 10: ...Copyright 1996 Baker Hughes Company viii ADTS 403 405 Calibration Manual English...

Страница 11: ...o verify performance following a main calibration Three options are available for main calibration and calibration check a Ps channel absolute calibration b Pt channel absolute calibration c Combined...

Страница 12: ...he unit datum level The datum level of the unit is on the front panel 1 6 1 Pressure Source The calibration menu prompts a selection of an internal or external pressure source The internal pressure so...

Страница 13: ...sensors This is particularly important when using an EXTERNAL pressure source The ADTS should be allowed one hour to thermally stabilize after power on and before performing any calibration routines...

Страница 14: ...Copyright 1996 Baker Hughes Company 4 ADTS 403 405 Calibration Manual English Chapter 1 Introduction...

Страница 15: ...NTROL Select ALT and enter a new aim point of 75 mbar abs Check that the ADTS attains the new aim point Make sure the pressure is stable and within 0 1 mbar of the aim point 10 Select RATE SPEED and L...

Страница 16: ...Copyright 1996 Baker Hughes Company 6 ADTS 403 405 Calibration Manual English Chapter 2 Leak Test...

Страница 17: ...Identification Number can be entered to prevent unauthorized entry into Calibration or Maintenance Mode To change the service PIN 1 Press 000 when the Main Calibration Menu is displayed 2 Enter the ne...

Страница 18: ...Copyright 1996 Baker Hughes Company 8 ADTS 403 405 Calibration Manual English Chapter 3 PIN Protection...

Страница 19: ...mple 4 The following screen then displays 5 Select the type of calibration to be carried out 6 Select the pressure source 7 Select the channel for calibration 8 Carry out the appropriate procedure in...

Страница 20: ...on standard to atmospheric pressure Disconnect calibration standard from the static output 8 If there is no further calibration or testing required switch off the unit 4 4 Calibration Check of the Pit...

Страница 21: ...nnect calibration standard to the pitot output 2 Adjust calibration standard to the first pressure value in Table 4 3 3 Compare the pressure value on calibration standard to the value displayed and re...

Страница 22: ...Rotate the screw beneath the label fully counter clockwise When calibration is complete rotate the screw fully clockwise and fit a new calibration label 4 6 2 Pressure Source Select the pressure and...

Страница 23: ...Pt channels to minimize calibration time and errors due to calibration standard non repeatability Connect the calibration standard to the pitot Pt output on the front panel Fit the blank on the Stati...

Страница 24: ...ing the rate selected by RATE If necessary check the current AIM value by select AIM Press QUIT to return to the calibration screen OFF This switches the pressure controllers OFF to enable leak testin...

Страница 25: ...ry out a leak test as detailed in Section 2 Set the unit to measurement mode main calibration external pressure source and Pt pitot calibration as detailed 4 1 1 Connect calibration standard to the pi...

Страница 26: ...itot range sensors Note A label on the front panel of the ADTS 403 405 shows the full scale range of the Pitot sensor 8 Press F4 END The unit calculates the change in zero FS and span RD from the prev...

Страница 27: ...sure and vacuum supplies need to be connected and the static Ps and pitot Pt outlets blanked To perform a rate calibration 1 Select RATE CAL and the unit automatically carries out a rate calibration T...

Страница 28: ...Copyright 1996 Baker Hughes Company 18 ADTS 403 405 Calibration Manual English Chapter 4 Calibration...

Страница 29: ...to return the current values of slope span offset zero and residual corrections curve fit The slope span offset zero values can then if necessary be adjusted and sent back and stored by the ADTS 403 4...

Страница 30: ...by offset adjustment 5 2 Defining Elements for a Program 5 2 1 Terms Variables and Constants The following terms variables and constants are used in this section The elements to be used in the program...

Страница 31: ...on point not being used 1 R n 0 2 NEXT N 5 2 3 To Find Pressure Errors Between Standard and ADTS 403 405 1 FOR n 1 to MAXPTS 2 Apply calibration pressure value from appropriate table 3 Store exact val...

Страница 32: ...0 M 1 MAXSLOPE AND ABS C MAXOFFSET THEN NEWSLOPE OLDSLOPE M NEWOFFSET OLDOFFSET C 5 2 4 3 To Calculate New Residual Correction Values 1 FOR n 1 to MAXPTS 2 3 4 5 NEXT n 6 R 0 0 all other missed residu...

Страница 33: ...libration Pressure inHg mbar 0 0 0 0 1 1 0 34 2 3 1 105 3 5 1 173 4 7 1 240 5 10 2 345 6 15 3 518 7 20 4 691 8 25 4 860 9 30 5 1033 10 34 5 1172 11 40 0 1355 Table 5 4 Pitot 850 knot Range Calibration...

Страница 34: ...lish Chapter 5 Calibration using IEEE 488 SCPI Table 5 5 Pitot 1000 knot Range Calibration Pressures Point Calibration Pressure inHg mbar 0 0 0 0 1 1 0 34 2 5 1 173 3 10 2 345 4 20 4 691 5 30 5 1033 6...

Страница 35: ...and calibrate Pt Reject Ps changes and return to calibrate Ps Complete Pt calibration Accept Pt changes and Quit Reject Pt changes and return to Pt calibration Go to new pressure set point Set Rate se...

Страница 36: ...tion END CAL data entry YES NO Enter Actual value Accept Pt changes and Quit Reject Pt changes and return to calibration menu Complete calibration Ps Pt abs ACTUAL VALUE data entry Complete Ps calibra...

Страница 37: ...ta entry ON OFF Go to new pressure set point Set Rate set point Turn controllers ON Turn controllers OFF Pt abs data entry AIM RATE data entry ON OFF Go to new pressure set point Set Rate set point Tu...

Страница 38: ...Copyright 1996 Baker Hughes Company 28 ADTS 403 405 Calibration Manual English Appendix A Quick Reference...

Страница 39: ......

Страница 40: ...kerhughes com Bangalore 91 9986024426 aneesh madhav bakerhughes com Milan 39 02 36 04 28 42 csd italia bakerhughes com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 3...

Отзывы: