OPERATING INSTRUCTIONS
DROPER FIELD 1000
FULLY MECHANICAL INFUSION PUMP
PROMOVET
sarl
ZAC de la Route de Beauraing, 1 08600 GIVET
(France)
info
@
promovet.fr
Страница 1: ...1 16 OPERATING INSTRUCTIONS DROPER FIELD 1000 FULLY MECHANICAL INFUSION PUMP PROMOVET sarl ZAC de la Route de Beauraing 1 08600 GIVET France info promovet fr...
Страница 2: ...iption of the Droper Field 1000 3 2 Using the Droper Field 1000 4 3 Operating issues 11 4 End of infusion 12 5 Cleaning Disinfection 13 6 Periodic inspection 15 7 Storage 15 8 Warranty conditions 15 9...
Страница 3: ...3 16 1 DESCRIPTION OF THE DROPER...
Страница 4: ...nvironmental conditions It has been designed using materials capable of operating at both negative and positive temperatures The infusion flow rate is entirely dependent on the viscosity of the infusi...
Страница 5: ...ng his her middle fingers of each hand upwards and inwards This movement will result in pivoting the closing spring clips inwards and releasing the cover 3 While holding the two open levers in positio...
Страница 6: ...ain height is reached with air being positioned at the top of the drip chamber The flow will be estimated by counting the number of drops in the first minute of infusion b Infusions in hostile environ...
Страница 7: ...the cover forwards until it enters into contact with the bulge at the surface of the infusion bag Stand directly above the device with outstretched arms and press down on the cover placing the cushio...
Страница 8: ...point 2 4 above ensuring that the drip chamber is positioned with the outlet pointing downwards Close the Droper Field open the clamp and the Luer Lock cap located at the end of the line and expel the...
Страница 9: ...oximate pressure of 100 mbar If placed at a higher level and according to the atmospheric pressure conditions the pressure shall be approximately increased in mbar by the distance in cm separating thi...
Страница 10: ...Close the line using the roller clamp or flow regulator by turning them to the OFF position or by means of any other suitable device To open the Droper Field 1000 during an infusion a higher amount o...
Страница 11: ...and re close the cover In the event that the cover is closed correctly open the Droper Field 1000 unfasten the mobile plate from its tightening clips check the following points and implement the relev...
Страница 12: ...ient or slightly below the level of the latter the net flow of the solute at the 3 way stopcock without blood reflux into the infusion line signals the absence of permeability at the infusion site The...
Страница 13: ...g clips releasing the plate from the mechanism see diagram While holding the Droper Field 1000 on the flat surface firmly grip the mechanism and pull it upwards This will result in the tightening clip...
Страница 14: ...the manufacturer s recommendations and at a distance of 30 cm from the device while ensuring that no liquid product is accumulated on the device For further information please contact your company s a...
Страница 15: ...he same height as the device and measure the time required to drain the bag This time must be between 320 and 340 seconds 7 STORAGE The device must be stored in a temperature controlled dry place Reco...
Страница 16: ...for a period of 10 years from its date of manufacture The user is responsible for recording the date of first use by any means deemed appropriate for traceability purposes 10 USEFUL CONTACTS Sales Dep...