Drive Medical Primus Medical PrimePlus P1752 Скачать руководство пользователя страница 25

 

 

 

 

P1752 Owner's Manual| REV 002 

 

April 21, 2015 

The information in this document is subject to change without notice.

 

 

Page 24 

CAUTION

 

 

If improperly performed, equipment or property DAMAGE or resident INJURY is 
possible during maintenance. 

 

 

The battery is not a toy for children to play with. 

 

Maintenance Notes: 

 

Following longer periods of storage, the battery may need to be charged 
and discharged several times so that performance reaches a peak. 

 

Batteries will lose their charge during storage.  Battery should be fully charged 
prior to storing.  A battery in storage should be recharged every 3 months at a 
minimum. 

Care tips 

 

Make sure the battery is kept clean and dry at all times. 

 

The battery must receive sufficient ventilation when charging. 

 

Remove backup battery when bed is not likely to be used for some time. 

Safety tips 

 

Don’t open or damage the battery. 

 

Don’t expose the battery to heat or open flames. Avoid storing it in direct 
sunlight. 

 

If the battery leaks and you come into contact with the fluid, wash the 
affected area thoroughly in water and consult a doctor immediately. 

 

Don’t throw the battery into fire, open it, solder or weld to it. 

Cleaning Notes 

 

Before cleaning, always unplug mains cable from wall outlet. 

 

When cleaning, take care not to damage the connection cable. 

 

Clean plastic surfaces such as the housing with slightly damp cloth. Never use 
cleaning agents on battery. 

 

Never use a high-pressure cleaner on the battery or spray fluids directly onto 
them using jets. You risk damaging the equipment. 

 

Solvents, alcohol, bleach, caustic agents, high acid or alkaline solutions or 
petroleum based products should not be used on the bed. 
 

NOTE:

 This bed NOT compatible with wash down tunnels. 

WARNING

 

Batteries must be disposed of in accordance with local regulations. 

 

 
 
 
 
 

Содержание Primus Medical PrimePlus P1752

Страница 1: ...cal com Bariatric 750 Full Electric Long Term Care Bed Model P1752 Owner s Manual This manual must be given to the Owner User of this bed and should be read carefully before putting this product into...

Страница 2: ...9 ELECTROMAGNETIC EMMISSION AND IMMUNITY 10 FEATURES AND ACCESSORIES 13 TECHNICAL SPECIFICATIONS 14 COMPLIANCE INFORMATION 15 ASSEMBLY 16 FREQUENTLY USED BED FUNCTIONS 26 BED MOBILIZATION AND STABILIZ...

Страница 3: ...personnel from possible hazards related to electrical shock FIRE HAZARD Information that appears under the FIRE HAZARD symbol concerns the protection of personnel from possible hazards related to fire...

Страница 4: ...foot deck position According to ISO 7000 1641 Operating instructions Change auto contour position Recycle the item in accordance with local regulations Change hi lo position of mattress support platfo...

Страница 5: ...ills to understand the information in the owner s manual operate the bed or control the movements of the bed through the control interfaces This bed should not be used or operated by children GENERAL...

Страница 6: ...to avoid squeezing those between parts of the medical bed DO NOT use an extension cord DO NOT allow the cord electrical outlets and electrical control box or hand pendant to become wet or submerged D...

Страница 7: ...tton when bed does not function as expected NEVER permit more than one 1 person in on the bed at any time NEVER exceed the weight capacity of this bed The weight capacity of this bed is 750lbs 340kg i...

Страница 8: ...or out of bed NEVER allow anyone under the bed at any time Supervision is required when this product is operated by or near children or people with disabilities DO NOT let any limbs or body parts pro...

Страница 9: ...rapment It also provides manufacturers with basic design criteria to consider when developing hospital convalescent beds recommends specific test methods to assess the conformance of existing hospital...

Страница 10: ...e mattress support platform must be used See Technical Specifications page for compatible mattress dimensions EMERGENCY EVACUATION If your mattress has Velcro side bolsters remove them and adjust matt...

Страница 11: ...UNITY of the EQUIPMENT or SYSTEM DECLARATION ELECTROMAGNETIC EMISSIONS Guidance and manufacturer s declaration Electromagnetic emissions The PrimePlus P1752 is intended for use in the electromagnetic...

Страница 12: ...e communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12...

Страница 13: ...in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in t...

Страница 14: ...ed Mounted Heavy Duty Casters Staff control panel Mattress Retainers Soft Gray Powder Coated Frame Color Wall Guard Assist Rails 80 or 84 Sleep Surface Bed End Panels WARNING Incompatible assist rails...

Страница 15: ...gh deck angle range 0 to 38 3 Foot deck angle range 0 to 22 3 Trendelenburg Reverse Trendelenburg capable No Mattress length 80 mattress deck length 80 203 cm Mattress length 84 mattress deck length 8...

Страница 16: ...0 75 30 75 Atmospheric pressure 86KPa 106Kpa 86KPa 106KPa This bed is classified for intermittent operation Bed actuators may be operated for 2 minutes out of every 20 minutes Applied Parts List Bed F...

Страница 17: ...240 VAC outlet Check that actuator cords are connected properly to the controller Raise the bed frame and check to ensure all cables are plugged securely into the control box and no wires are loose V...

Страница 18: ...the sides of the bed in the appropriate holes for the width mattress being used Width options are detailed in the technical specification section HEADBOARD AND FOOTBOARD INSTALLATION AND ADJUSTMENT T...

Страница 19: ...adboard footboard down until it stops Screw the knobs back in on each side of the support assembly to lock the headboard footboard in position ASSIST RAIL ASSEMBLY TOOLS REQUIRED Wrench Socket 5 16 He...

Страница 20: ...assist rail ASSIST RAIL WARNING Other manufacturers assist rails may not be compatible and can lead to entrapment issues or harm to patients residents Only compatible Primus Medical by Drive assist r...

Страница 21: ...7 mobility position only 39 42 and 48 The width extension expands by using the push button mechanism Push the expansion button mechanism in and pull out width extension until you reach the desired wid...

Страница 22: ...board support bracket Repeat process with the footboard support bracket The end mattress retainers must be properly adjusted for the length of the mattress being used NOTE A Gap Fill may also be used...

Страница 23: ...e Wall Guard into the holes of the mount underneath the bed Place pins on each side of the bar into the holes to keep the Wall Guard in place The Wall Guard is now in place BATTERY BACK UP TOOLS REQUI...

Страница 24: ...MATION WARNING The battery backup should be used for EMERGENCY PURPOSES ONLY The only compatible battery backup is a DewertAG7 battery DO NOT use the battery backup to demonstrate or display a bed If...

Страница 25: ...to be used for some time Safety tips Don t open or damage the battery Don t expose the battery to heat or open flames Avoid storing it in direct sunlight If the battery leaks and you come into contact...

Страница 26: ...y 4 3 8 Hex bolts each Use 2 3 8 bolts to attach the Cross Bar C to both vertical Trapeze arms at one end of the bed and then use 2 3 8 bolts to attach the cross bar to both vertical Trapeze arms at t...

Страница 27: ...decreased mental acuity should NOT have access to pendant NOTE For every 2 minutes of continuous actuator use the bed actuators must not be used for 18 minutes 4 Function Staff Control Pad Operation o...

Страница 28: ...lock the bed frame lift up all 4 caster locks with your foot CAUTION Moving the bed while the floor lock or caster lock is engaged may cause DAMAGE to the bed or floor DO NOT move the bed unless the f...

Страница 29: ...wash or steam clean any parts Rinse completely with water Maximum temperature 110oF or 43oC DO NOT rinse with cold water Ensure all parts are dry before using or storing NOTE This bed is NOT compatibl...

Страница 30: ...or tighten as necessary Ensure welds DO NOT have cracks and visually inspect for stress Ensure no tubes are bent If the bed has a battery unplug bed from wall outlet and check for proper function Che...

Страница 31: ...ose tighten and if missing replace Lubricate clevis pins at hinge points Lubricate Hi Lo tracks of bed for smooth travel Actuators Inspect push tubes and end connections of all actuators for excess we...

Страница 32: ...place by 2 clevis pins Remove bowtie clips from clevis pins Slide the clevis pins out of the holes Replace the actuator To reassemble bed reverse previous steps and make sure to o Assemble clevis pins...

Страница 33: ...should be done by a skilled technician Effect Possible Cause Verification Corrective Action Bed does not stay in place Caster lock is not engaged Caster locks not in locked position Activate caster Lo...

Страница 34: ...alls while operating Thermal shut down Bed works for a short period of time then cuts out Check for obstructions or any interference with bed frame such as window seal or too much weight on the bed fr...

Страница 35: ...cleaning repair accident negligence unauthorized alteration or use in inappropriate environmental conditions or failure to maintain the product consistent with user and service instructions Any change...

Страница 36: ...the sleep deck on the frame rail below the mattress support platform on either side of the bed Sales and Service Contact Drive Medical if you need any help with installation usage or maintenance of t...

Страница 37: ...ain uncontrolled body movement or who get out of bed and walk unsafely without assistance must be carefully assessed for the best ways to keep them from harm such as falling Assessment by the patient...

Страница 38: ...tions and pain relief When bed rails are used perform an on going assessment of the patient s physical and mental status closely monitor high risk patients Consider the following Lower one or more sec...

Страница 39: ...on American Medical Directors Association American Nurses Association American Society for Healthcare Engineering of the American Hospital Association American Society for Healthcare Risk Management B...

Страница 40: ...REV 002 04 21 2015 April 21 2015 Page 39 The information in this document is subject to change without notice Drive Medical 99 Seaview Blvd Port Washington NY 11050 Ph 1 877 224 0946 Fax 516 988 4601...

Отзывы: