P1752 Owner's Manual| REV 002.04.21.2015
April 21, 2015
Page 23
The information in this document is subject to change without notice.
Slide the battery into place; make sure it is positioned securely into the
mounting bracket.
Attach the cord from the battery into the control box (black color).
The battery back-up will now work if the bed becomes unplugged from the
wall outlet or there is a power failure.
NOTE:
Allow the battery to charge for a minimum of 24 hours before first use
Battery back-up is specified as a part of bed
Battery is not to be used in a damp environment such as outdoors
BATTERY BACK-UP INFORMATION
WARNING
The battery backup should be used for EMERGENCY PURPOSES ONLY. The
only compatible battery backup is a DewertAG7 battery.
DO NOT use the battery backup to demonstrate or display a bed.
If the battery backup is used, it MUST be recharged before becoming drained
and dormant. If the battery is run dead and left for any extended period of
time, the battery can go into HIBERNATION and cannot recover or be
recharged.
Battery function should be checked periodically to insure battery health.
Primus Medical by Drive cannot be held responsible for improper backup use.
When a battery is connected to the control box, unplugging the bed from
mains power will not switch bed off.
The battery must be replaced if the housing or cable is damaged.
The battery must be replaced if there is leakage from the battery.