P1752 Owner's Manual| REV 002
April 21, 2015
The information in this document is subject to change without notice.
Page 16
Remove the mattress retainers from the bed frame. They are wrapped in
foam and zip tied to the frame.
Remove any remaining zip ties or foam left on bed frame.
Make sure all connections to the control box are secure
The product should be positioned in a way that the mains plug is easily
plugged into and unplugged from a wall outlet
NOTE:
DO NOT remove zip ties that are holding cords underneath bed frame.
INITIAL INSPECTION
Inspection of All Components – Receipt of assembled bed
Check bed components for obvious damage.
Inspect power supply cords for cuts and/or damage
If power cord is free of damage, plug into grounded 110-240 VAC outlet
Check that actuator cords are connected properly to the controller.
Raise the bed frame and check to ensure all cables are plugged securely
into the control box and no wires are loose.
Verify proper functionality of all features
AFTER ASSEMBLY
.
ASSEMBLY
TOOLS REQUIRED
¾” Wrench and Socket
5/32" Hex Key
5/16” Hex Key
(2) 3/16” Hex Key
KNOWLEDGE/SKILLS REQUIRED
Ability to read and understand
owner’s manual
Ability to manipulate required
tools
Basic mechanical aptitude
MATTRESS RETAINER INSTALLATION/ADJUSTMENT
Mattress retainers are designed to keep the mattress in place on the sleep
surface.
Determine size of your mattress
There are two different sizes of mattress retainers. 4 narrow retainers for the
sides of the bed and 2 wider retainers for the end of the bed.
Insert the two end mattress retainers in the holes appropriate for the length
mattress being used. Length options are detailed in the technical