Guidance and manufacturer's declaration: Electromagnetic Immunity
The PrimePlus™P1752 is intended for use in the electromagnetic environment specified below. The customer or the
user of the PrimePlus™P1752 should ensure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conducted
RF
IEC 61000-4-6
Radiated RF
IEC
61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to
any part of the PrimePlus™P1752 including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
80MHz to 800MHz
800MHz to 2,5MHz
Where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the Transmitter
manufacturer and
d
is the recommended
separation distance in meters (m). Field
strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey,
a
should be less than the
compliance level in each frequency
range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE:
I At 80MHz and 800MHz the higher frequency range applies.
NOTE:
2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which thePrimePlus™P1752 is
used exceeds the applicable RF compliance level above, the PrimePlus™P1752 should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the PrimePlus™P1752.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
[V
i
]
V/m.