2. For those with a substantial amount of staining dilution drug
(such as methylene blue, indigo green and acid indigo blue),
or carbon monoxide hemoglobin (COHb), or methionine
(Me+Hb) or thiosalicylic hemoglobin, and some with icterus
problem, the SpO
2
determination by this monitor may be
inaccurate.
3. The drugs like dopamine, procaine, prilocaine, lidocaine and
butacaine may also be a major factor blamed for serious error
of SpO
2
measure.
4. As the SpO
2
value serves as a reference value for judgment of
anemic anoxia and toxic anoxia, some patients with serious
anemia may also report good SpO
2
measurement.