The conserver must:
Be used only after one or more settings have
been individually determined or prescribed for
you at your specific activity levels.
Be used with a specific combination of parts and
accessories (i.e. cannulas, tubing) that meet the
specifications of the conserver manufacturer,
and that were used while your settings were
determined.
Your prescribed settings should be periodically
reassessed by a physician or other appropriately
licensed healthcare practitioner for effectiveness of
therapy.
WARNING (Cont.)
Use only parts and accessories recommended by
the manufacturer to ensure proper function and to
avoid the risk of fire and burns.
It is unsafe to use accessories, detachable parts and
materials not described in the instructions for use,
interconnect this equipment with other equipment
not described in the instructions or modify the
equipment.
Cannula tubing can pose a strangulation risk and
tripping hazard. Locate tubing away from areas of
high foot traffic and keep children and pets away
from cannula tubing.
Always maintain a backup supply of oxygen (i.e.
compressed cylinders and regulator).
•
•
•
•
•
•
•
•
•
Not intended for use with any humidification devices
.
•
Do not use cannula tubing that is longer than 7 ft.
(2.13 m).
Not for use with an endotracheal tube or
tracheostomy
.
•
Lint and dust can contaminate the oxygen path and
create a fire hazard. Keep all oxygen connections clean
and free from lint and dust.
Make sure the fill valve, if equipped, is clean and free
of dirt, debris, oil or other contaminants.
To reduce the risk of rapid oxygen recompression and
fire, open the cylinder valve slowly and completely so
the pressure gauge moves slowly as it indicates the
cylinder pressure.
Do not submerge the conserver in the disinfectant,
wipe disinfect only.
•
•
•
•
•
•
•
•
•
•
Close supervision is necessary when the nasal
cannula
is
used
by
impaired
persons
or
near children
.
WARNING (Cont.)
The proper placement of the nasal cannula in the nose
is critical to the consistent operation of this equipment.
Do not wet the area of the oxygen inlet connection,
the disinfectant residue may create a risk of fire.
Before using this unit, make certain that your hands
are free of oil, grease and other contaminants,
and that the
cylinder is secure in an upright position.
•
Do not attempt to open or remove enclosure.
There are no user-serviceable internal components.
Choking Hazard - The 870 post seal washer may pose
a choking hazard. Small parts not for children under 3
years or any individuals who have a tendency to
place inedible objects in their mouths.
r models or brands of oxygen
The settings of othe
therapy equipment do not correspond with the
settings of the OM-812CE Pneumatic Oxygen
Conserver.
5
Содержание Bonsai OM-812CE
Страница 21: ...Pneumatic Diagram Pneumatic Diagram depicts device in conserving selection 21...
Страница 122: ......
Страница 123: ......