TROUBLESHOOTING
PROBLEM
PROBABLE CAUSE
SOLUTION
NOTE:
Do not attempt to open the unit. If the unit is opened or tampered with, the warranty is void.
Unit does not pulse.
No
oxygen
delivery.
Unit pulses or
flows
continuously.
Cylinder valve is closed.
Turn
cylinder
valve counter
clockwise
to
open.
Cylinder is empty.
Check the oxygen gauge. Replace the cylinder,
if empty.
Oxygen cannula is blocked or kinked.
Remove kinks. Clean or replace, if necessary.
Selector switch is set to the “OFF” position.
Make sure the selector switch is set to the
appropriate delivery setting.
Unit is set to the “cf” position.
Turn the selector switch to the appropriate
delivery setting.
Unit was not set to “OFF” prior to opening
the cylinder valve.
Turn the selector switch to “OFF,” wait a few
moments, then set at proper delivery setting.
Vent hole is obstructed.
Remove obstructions, such as labels or a tight-
fitting carrying bag, and resume use as usual.
Fault in the unit.
Continue the therapy by setting the selector
switch to
“cf” (continuous flow). This setting
increases the oxygen consumption, so you
should regularly check how much oxygen
you have left.
Non-functioning units are subject to warranty provisions and the manufacturer repair/return policy.
If necessary, call your Home Care Provider.