5. Mode Control

Expose the controls by sliding front cover down from top of unit. This switch

has 3 positions: B for Burst stimulation, N for Constant stimulation, and M for

modulation stimulation. Push the Mode Selector until engaged in position

desired.

6. Pulse Rate Control:

This dial determines how many electrical impulses are applied through the skin

each second. By turning these controls, the number of current impulses per

second(Hz) for both channels can be continually adjusted. Unless otherwise

instructed, turn the pulse rate control to the 70-120 Hz range.

7. Pulse Width Control:

This dial adjusts the length of time each electrical signal is applied through

the skin, which controls the strength and sensation of the stimulation. If no

instructions regarding the pulse width are given in therapy, set the control to

the suggested 70-120 uS setting.

higher frequencies, the lamp will appear to be constantly illuminated.

3. On/Off Switch and Intensity Control :

If both controls are in the off-position (white markings on the housing), the

device is switched off.

By turning the controls clockwise, the appropriate channel is switched on

and the impulse display led will illuminate and begin to pulse according to

the frequency set.

The current strength of the impulses transmitted to the electrodes increases

further when the control is turned clockwise.

To reduce the current strength or switch the device off , turn the controls counter

clockwise  to the required setting or off- position.

4. Lead Connector

Connection of the electrodes is made with two-lead connector. The device

must be switched off before connecting the cables. Both intensity controls

must be at the Off position. Electrodes must be pressed firmly on the skin.

PDF created with FinePrint pdfFactory trial version 

http://www.pdffactory.com

Содержание AGF-3X

Страница 1: ...Port Washington NY 11050 PDF created with FinePrint pdfFactory trial version http www pdffactory com...

Страница 2: ...ead Wires 10 12 Lead Wire Maintenance 10 13 Electrode Options 10 14 Electrode Placement 11 15 Tips For Skin Care 11 16 Application of Re usable Self Adhesive Electrodes 12 17 Adjusting the Controls 13...

Страница 3: ...OF PAIN Pain is a warning system and the body s method of telling us that something is wrong Pain is important without it abnormal conditions may go undetected causing damage or injury to vital parts...

Страница 4: ...hat when operating potentially dangerous ma chinery the stimulator controls are not changed abruptly 11 Electrodes should not be placed over the eyes in the mouth or internally 12 Keep this device out...

Страница 5: ...hannels 2 Pulse Ampulitude Adjustable 0 80 mA peak into 500 ohm load each channel 3 Pulse Rate Adjustable from 2 to 150 Hz 4 Pulse Width Adjustable from 30 to 260 microseconds 5 Modulation Mode Pulse...

Страница 6: ...pain relief should occur shortly after the intensity setting has been determined However in some cases pain relief may take as long as 30 minutes to achieve TENS units are typically operated for long...

Страница 7: ...od allowing the current flow on either side of the painful area generally through the nerve branches that feed into the pain site Chapter 15 TIPS FOR SKIN CARE To avoid skin irritation especiallyifyou...

Страница 8: ...d across the skin during application To prevent this apply electrodes from centre outward avoid stretching over the skin 5 To minimize pulling stress tape extra lengths of lead wires to the skin in a...

Страница 9: ...rapy set the control to the suggested 70 120 uS setting higher frequencies the lamp will appear to be constantly illuminated 3 On Off Switch and Intensity Control If both controls are in the off posit...

Страница 10: ...work with open lights for example candles etc when working with flammable liquids 2 Stains and spots can be removed with a cleaning agent 3 Do not submerge the device in liquids or expose it to large...

Страница 11: ...nd accessories in carrying box and keep it in cool dry place 6 The packed TENS device should be stored and transported under the temperature range of 20 60 relative humidity 20 95 atmo sphere pressure...

Страница 12: ...PDF created with FinePrint pdfFactory trial version http www pdffactory com...

Отзывы: