HPG 13150 400 V (A)
Communication Interfaces
4-5
HPG™ 13150 400 V RF Power Generator
Prerelease version
The HPG Generator may be shipped with an interlock jumper plug that
provides a connection between the interlock pins. You can use this jumper
plug to satisfy the interlock and enable operation in situations where you do
not intend to connect the remaining pins on this port.
25-PIN USER PORT CABLING REQUIREMENTS
Connect the HPG generator’s 25-pin
User Port
to the system controller with a
shielded, 25-wire I/O cable. Shielded twisted-pair wiring is recommended but
not mandatory. Minimize signal losses by keeping the cable as short as
possible. The maximum recommended cable length between the HPG
generator and the controller is 10 meters (33 feet). Minimize interference from
adjacent electrical equipment by terminating the EMI shield in the cable to the
metal shells of the cable’s connectors. Additionally, you must tie the chassis of
the HPG generator to a local earth ground through an adequately sized
copper grounding strap.
Note:
Grounding the
User Port
at the HPG generator reduces noise
interference. To avoid ground loop problems, you should typically
ground only one end of the
User Port
cable.
ACTIVATING THE 25-PIN USER PORT
To activate the 25-pin User Port control
1. Press the
Program
soft key.
2. Press
Next
until you see
Device Configuration
on the display.
3. Press
Change
.
4. Press
OK
until you see
Control by
on the right-hand side of the screen.
5. Press
User Port
.
6. Press
Execute
.
Using the interlock jumper plug disables the interlock function.
Содержание 13150 400 V
Страница 2: ......
Страница 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Страница 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 24: ...1 10 Safety and Product Compliance Guidelines HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Страница 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...