Table 7: Temperature Measurement Accuracy Range
Temperature Range
Measurement Accuracy
-40°F to +32°F (-40°C to 0°C)
+/- 0.4°F (+/- 0.2°C)
32°F to 140°F (0°C to 60°C)
+/- 0.5°F (+/- 0.3°C)
You can configure the units of measure (°F or °C) and the reference base temperature by using the
Dresser MeterWare software.
For ease of calibration, you can use the Dresser MeterWare software to perform a single point
temperature field calibration.
For more information, refer to the
Dresser MeterWare Software Manual
.
11.4
Flow Rate Measurement
Access the flow rate by scrolling to the Flow Rate screen (
FLOWRATE
) on the meter’s LCD screen
, as
described in Section 4.1. The flow rate indication is not instantaneous, but it is based on the average flow
rate of the last thirty (30) seconds of the saved uncorrected volume data.
Because the meter updates and stores the data collected every thirty (30) seconds, there is a slight delay
in the timing of the displayed results ranging from 1 to 29 seconds until the results are updated again.
11.5
Faults and Alarms
11.5.1
Fault Types
A fault indicates a problem with the meter electronics hardware or the firmware.
Table 8: Fault Types
Fault Type
LCD Screen
Display
Occurs When
Temperature Fault
T FLT
The temperature probe is faulty or disconnected from the meter
electronics.
Volume Fault
VOL FLT
The volume input pulses are out of sequence.
Volume Sensor Fault
VOL SNSR FLT
An open wire on the volume input board or mag pickup is
present.
Internal Operation Fault
INT FLT
A memory access fault is present.
Power Fault
PWR FLT
A power problem occurred or battery wiring is disconnected.
Low Battery Fault*
LOW BATT
The battery voltage drops below 2.7 V.
* Refer to Section 11.5.5.