5. Glass pieces should be handled with care when installing, all the joints must be coated
with vacuum grease.
6. Can not be used under the environment full of flammable, explosive and narcotic gas.

IV. Main technical parameters

Capacity of rotation bottle

5L

Heating power of water bath

2000VA

Collection bottle capacity

3L bottom discharge Fluctuation range of water bath: 180mm

Rotation speed:10~110r/min Frequency control of motor speed

Power

~220v/50Hz

Rotating motor

120VA AC induction

Dimensions

950×500×1200mm

Weight (net)

50kg

V. Installation steps

1. After unpacking, check according to the packing list, and assemble according to Figure
I after scrubbing and drying them.
2. Principal machine should be aligned with water tank. If the ground surface is not flat,
smooth down by using hard objects.
3.

Install the four-way valve

: connect the valve to the principal machine, encase the

forcing screw into four-way flange mouth, encase the nylon collar, align the internal jointing
washer, and screw the forcing screw tightly.
4.

Install the condensation pipe

: insert it into the four-way ball-milling position and keep

vertical, screw the fixed packet ring with the slotted screwdriver tightly, and then screw the
cross clip tightly. Fix the condensation pipe with four-way butt joint through a special
connector.
5.

Install the condensation elbow

: connect the concave ball milling of condensation

elbow with the four-way ball milling, and clamp with the ball milling mouth clip; connect the
convex ball milling of condensation elbow with the connection bottle, and install the ball
milling mouth clip.
6. Insert the deeding tube into the standard concave milling mouth of four-way valve.
7. Adjust the position of all brackets, tight degree of screws, and fitness degree of glass
joints.

Содержание ZNHW-500ML

Страница 1: ... and operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Страница 2: ...Rotary Evaporator Specification Thank you for choosing XTRACTION SERVICES XD 5000 rotary evaporator our mission make your research more accurate and efficient ...

Страница 3: ...e of the device and reduce its performance Therefore please be sure to carefully read the manual before using The manual will be submitted to end users Please keep this manual to prepare for inspection when required The manual is used for only reference the specific shape design is subject to the real object I Usage XD 5000 production type rotary evaporator is a kind of device with the operation p...

Страница 4: ...uninstallation 5 Thermostat adopts the touchable smart digital display which can accurately set the heating temperature 6 Frequency control is available for the adjustment of rotation speed with stable and reliable digital display 7 Water bath kettle can be moved up and down The XD 5000 type kettle adopts the method of single hand wheel lift and XD 5000 adopts two ways namely through power and han...

Страница 5: ...hard objects 3 Install the four way valve connect the valve to the principal machine encase the forcing screw into four way flange mouth encase the nylon collar align the internal jointing washer and screw the forcing screw tightly 4 Install the condensation pipe insert it into the four way ball milling position and keep vertical screw the fixed packet ring with the slotted screwdriver tightly and...

Страница 6: ... of vacuum degree 11 Only after being injected with water can the water bath be powered and heated Operation a Wiring and electricity charging must be done by professional electricians and ground wire terminal must be grounded Power utility power 220v 50Hz b Press the power switch to the state of ON and turn the speed control knob when the principal machine will begin to rotate c After injecting w...

Страница 7: ...tly with hands Water tank can not be heated 1 Check the 15A fuse 2 Check whether the heating plug is loose 1 Replace the 15A fuse 2 Insert the heating plug tightly Insufficient vacuum 1 Check whether the seal ring is worn 2 Check the performance of vacuum pump and whether each connector is tightened block the air outlet of vacuum pump to observe the indicators on vacuum table 3 Check whether the g...

Страница 8: ...he warranty period for this product is one and half a years commencing from the leaving the factory after being packed well or one year after users purchase Under the premise of using it according to the norms for storage and usage if the product can not work normally we will provide the service of repair or replacement free of charge Glass pieces are beyond the scope of warranty In case of improv...

Страница 9: ...tube 1 set 8 Stainless steel ball milling mouth clip 35 2 9 Stainless steel rod bracket 1 10 German cross clip 1 11 Aluminum cross clip Black 1 12 Fixed packet ring 1 13 Bottle supporting rack 1 14 Fuse 20A 2 each 15 Cross screwdriver 1 16 Flask lock wrench 1 17 Certificate warranty card 1 set Product certification Product name rotary evaporator Product model XD 5000 Production date Product number...

Страница 10: ...Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Отзывы: