Drawell SMART-e Series Скачать руководство пользователя страница 14

-

12

-

Trouble

Causation

Remedy

Can’t open software.

Plug soft dog again.

Check soft dog.

Can’t browse image.

The data link or damage not, Driver

files destroyed.

Check whether good or replace

data interpolation, To install the

driver.

Blank screen

Microscope untapped or equipment

photosensitive

brightness

not

enough.

Added microscope brightness or

equipment exposure value

Image corrugated

flashing

Brightness too high.

Reduce

brightness

and

appropriate

supplementary

equipment microscope exposure

value.

Browse the image

delay

computer

system

burden

or

hardware configuration is low.

Clean

or

replace

computer

computer system.

Image is not clear

No focusing, slice in improper

position.

Adjusting fine focusing knob;

moving slice’s position or replace

slice.

Images are not

synchronous

Diopter adjusting position deviation

Adjusting diopter adjusting ring.

Software calibration

error

Incorrect operation procedure

Please carefully reference manual

procedure.

Calibration select error

Among the image and to gauge

calibration of multiple objective to

correction.

When finding trouble , at first ,please confirm whether microscope in good condition, then further

testing problems in digital system. If unable to dispose, please contact the agency or manufacturer

to solve the problem with specific guidance.

Appendix 2: Digital System

Technical Parameter

Model

SM AR T

e

-3 2 0

SM A R T

e

-5 0 0

Image Collector
Transfer

3.2MP 1/2 " CMOS

Color progressive scan

imaging sensor

5MP 1/2 " CMOS

Color progressive scan

imaging sensor

Resolution

2048×1536 Pixel

maximum

2592×1944 Pixel

maximum

Frame Rate

15

30 fps

15

30 fps

Optical Camera
Lens

0.5×/1×

Data Transmission

USB2.0 connection

Transmission

Rate 480mps

without external

power supply.

Operating System

Support WINDOWS2000 / XP / Vista System

Image Analysis
System

OPTPro2008 Digital Image Analysis System

Содержание SMART-e Series

Страница 1: ...operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 008...

Страница 2: ...SMART e Series Digital Microscope Operation Instruction To ensure the safety and obtain satisfactory performance please study this operation instruction thoroughly before your operation...

Страница 3: ...And Carry 4 4 Operation And Use 5 5 Image System Operation 8 6 Maintenance 9 Appendix1 Troubleshooting 11 Appendix2 Digital System Technical Parameter 12 Appendix3 Standard Outfits 13 Please read the...

Страница 4: ...Screw or knurled thumb screw H Mechanical Stage Moving vertically and horizontally Knob J Right Coarse Fine Focusing Knobs K Brightness control knob F Body handle L Power switch M Condenser Focusing K...

Страница 5: ...nal Fine division 2 m Condenser up down range 26mm 7 Condenser Abbe condenser N A 1 25 Adjustable aperture Aperture center can be adjustable 8 Illumination Non spherical illumination System Field Diap...

Страница 6: ...n condenser screw raising condenser installing ring to the top Electrical Equipment Checking Plug power line in power input turn on power switch L observe brightness of light collector then continue t...

Страница 7: ...comfortable sensation of the coarse and fine focusing knob prevent moving downward Clockwise can relax instead can lock 5 Adjust condenser aperture diaphragm Adjust condenser focusing knob M and aper...

Страница 8: ...Rotate condenser focusing knob until in the middle of the field 3 Put 40 Objective in optical path rotate E will make field image slightly bigger than field Please readjust if field image isn t in mid...

Страница 9: ...to pull out power wire from charging power socket in order to prevent the backflow of discharge battery electric current 3 Charging machine please replace 3 batterys in same size when battery can t c...

Страница 10: ...o browse image collector function clear or basic clear image should also be on the screen if isn t satisfied only slightly rotating J N micro focusing hand wheel until get satisfactory imaging results...

Страница 11: ...and in carrying Don t drag it on the surface of worktable to avoid damage to microscope and worktable 6 3 Replacement of bulb 6 3 1 Replacement of halogen Please use specified bulbs If use other ligh...

Страница 12: ...t cut off power of microscope and pull out the power plug 2 Take off fuse base in the back of base take out old fuse 3 Replace a new fuse with same size then plug fuse base in 6 5 When not in use 1 Pl...

Страница 13: ...ace specimen correctly Surface of objective lens Is dirty especially it is easy for the front lens of 40 objective to dip in immersion oil Clean it Immersion oil isn t used for 100 objective oil Use i...

Страница 14: ...Adjusting diopter adjusting ring Software calibration error Incorrect operation procedure Please carefully reference manual procedure Calibration select error Among the image and to gauge calibration...

Страница 15: ...0 Rotatable 360 Interpupilary Distance 50 75mm Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76mm 52mm Scale 0 1mm Two Slide Holder Condenser Abbe N A 1 25 Iris Diaphragm Illuminatio...

Страница 16: ...ology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn...

Отзывы: