5 . Image System Operation

This product is equipped with various camera device, such as CMOS digital cameras, CCD camera and
VGA camera device etc, corresponding image output devices using the computer, LCD. And can choose by
0.5X or 1X optics cameras.

(1)

Adaptor for Digital Camera

To CMOS digital cameras, for example, remove USB2.0 connection lines, insert the square plug in USB2.0

10

behind the Integrative trinocular head, insert another flat plug in USB2.0 , make sure solid and stable.

(2)

Use Of Camera Device

After the connection, open image display equipment (such as computer and start image collector & analysis
software). Adjust microscope, adjust the eyepiece until clear image. Then open software to browse image
collector function, clear or basic clear image should also be on the screen (if isn’t satisfied only slightly
rotating J

N micro focusing hand wheel

, until get satisfactory imaging results.

Then can browse through the image acquisition software mode, or take images from the screen directly
observed

The product has passed factory inspection that eyepiece observation and photographic camera device
adopt with synchronous size. General adjustment synchronous with eyepiece fixed diopter (only for
model with single diopter adjusting ), but also result in slight error because the difference of the vision,
can be adjusted depending on eyepiece diopter adjusting ring as benchmark. With double depending on
the degree of adjustment, please circle one "0" scale line for the datum is conducted on the basis of
regulating use.

Through

USB

transmittocomputer.

Picture

Integrative
Trinocular

Head

Содержание SMART-e Series

Страница 1: ...operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 008...

Страница 2: ...SMART e Series Digital Microscope Operation Instruction To ensure the safety and obtain satisfactory performance please study this operation instruction thoroughly before your operation...

Страница 3: ...And Carry 4 4 Operation And Use 5 5 Image System Operation 8 6 Maintenance 9 Appendix1 Troubleshooting 11 Appendix2 Digital System Technical Parameter 12 Appendix3 Standard Outfits 13 Please read the...

Страница 4: ...Screw or knurled thumb screw H Mechanical Stage Moving vertically and horizontally Knob J Right Coarse Fine Focusing Knobs K Brightness control knob F Body handle L Power switch M Condenser Focusing K...

Страница 5: ...nal Fine division 2 m Condenser up down range 26mm 7 Condenser Abbe condenser N A 1 25 Adjustable aperture Aperture center can be adjustable 8 Illumination Non spherical illumination System Field Diap...

Страница 6: ...n condenser screw raising condenser installing ring to the top Electrical Equipment Checking Plug power line in power input turn on power switch L observe brightness of light collector then continue t...

Страница 7: ...comfortable sensation of the coarse and fine focusing knob prevent moving downward Clockwise can relax instead can lock 5 Adjust condenser aperture diaphragm Adjust condenser focusing knob M and aper...

Страница 8: ...Rotate condenser focusing knob until in the middle of the field 3 Put 40 Objective in optical path rotate E will make field image slightly bigger than field Please readjust if field image isn t in mid...

Страница 9: ...to pull out power wire from charging power socket in order to prevent the backflow of discharge battery electric current 3 Charging machine please replace 3 batterys in same size when battery can t c...

Страница 10: ...o browse image collector function clear or basic clear image should also be on the screen if isn t satisfied only slightly rotating J N micro focusing hand wheel until get satisfactory imaging results...

Страница 11: ...and in carrying Don t drag it on the surface of worktable to avoid damage to microscope and worktable 6 3 Replacement of bulb 6 3 1 Replacement of halogen Please use specified bulbs If use other ligh...

Страница 12: ...t cut off power of microscope and pull out the power plug 2 Take off fuse base in the back of base take out old fuse 3 Replace a new fuse with same size then plug fuse base in 6 5 When not in use 1 Pl...

Страница 13: ...ace specimen correctly Surface of objective lens Is dirty especially it is easy for the front lens of 40 objective to dip in immersion oil Clean it Immersion oil isn t used for 100 objective oil Use i...

Страница 14: ...Adjusting diopter adjusting ring Software calibration error Incorrect operation procedure Please carefully reference manual procedure Calibration select error Among the image and to gauge calibration...

Страница 15: ...0 Rotatable 360 Interpupilary Distance 50 75mm Nosepiece Quadplex Mechanical Stage Stage Size 145X140mm Travel 76mm 52mm Scale 0 1mm Two Slide Holder Condenser Abbe N A 1 25 Iris Diaphragm Illuminatio...

Страница 16: ...ology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd Greeland Max Mall PuDong Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn...

Отзывы: