Actual required sterilizing time (53min)=Delay time(33min) + Normal sterilizing time(20min)

FD50A Sterilizing Delay Time Reference Value (Every Flask)

Liquid Volume

Delay Time

3L

30min

2L

25min

1L

20min

500cc

15min

Never press the "STOP"button casually to stop the running program during sterilizing of

liquids to avoid the liquid overflowing into the sterilizer chamber and even blocking the
valve and pipe lines.

7. Starting sterilization

After setting the sterilization parameter, press "START" button to start work, "STER"
indicator will start to blink, when reaching the setting sterilization time, the sterilizing time
on the screen start to count down.

Note:

In order to prevent misuse, "START" and "STOP" button have a delayed response

function. Keep the button for 2 seconds to start work.

8. Opening the cover & Program finished

1) When the set sterilizing time is up, the system will send out a sound alarm, and the
steam will be exhausted to release automatically "EXHT" indicator starts to blink.
2) Under solid mode with drying, when the temperature is lower than the boiling point, the
system will send out a sound alarm, and the screen will show "OPEN" to indicate the
opening of the cover. After opening the cover, system will start drying function, "DRY"
indicator starts to blink, drying time on the screen starts to count down.

3) When program finished and the temperature is below 97

C

(Solid mode or solid+drying mode) or below 80

°C

(liquid mode),

the "COMP" indicator will blink, which means the sterilization is finished.

9.Taking the sterilized articles out

1)Always wear the heat insulation gloves when taking articles after sterilizing out of the sterilizer chamber, and wait until the
steam disperses before reaching into the sterilizer chamber.
2)When sterilizing the liquids, verify the temperature is low enough before taking it out to avoid scald due to low cooling

Содержание DW-FD Sereries

Страница 1: ...Drawell Scientific USER S MANUAL DW FD Series FD Series Automatic Autoclave...

Страница 2: ...oading articles to be sterilized 08 5 Closing the sterilizer chamber cover 09 6 Selecting sterilizing program 09 7 Starting sterilization 10 8 Opening the cover Program finished 10 9 Taking the steril...

Страница 3: ...is OMpa before opening the chamber cover Never attempt to open the chamber cover and drainage valve or draw out the steam trapping bottle when the pressure inside the sterilizer chamber is over OMpa...

Страница 4: ...peration to allow the instrument to cool down so that it will be able to produce adequate saturated steam In case of abnormal conditions such as noise odor and smog please turn off the power immediate...

Страница 5: ...Drawell Scientific Chapter 1 External Features Controller I External Features...

Страница 6: ...Drawell Scientific...

Страница 7: ...sure gauge 0 0 4Mpa Working environment 5 40 C relative humidity 10 85 Rated working pressure 0 25Mpa Designed pressure 0 3Mpa Sterilization temperature 105 135 Sterilization timer 1 300min Drying tim...

Страница 8: ...ake out the water plate remove the packaging place it in the bottom of the chamber in a horizontal position Connectioninstrument must be grounded reliably if the power socket does not 1 The instrument...

Страница 9: ...to avoid affecting normal work of sterilization Caution If the elevation is not set properly machine can not work normally Chapter 4 Operation Instruction I Basic Sterilization Illustration Operation...

Страница 10: ...amber cover too fiercely so that it will not be extruded and broken 4 Loading articles to be sterilized 1 Put the articles to be sterilized into the baskets and close the cover 2 Make sure the water p...

Страница 11: ...icator Locked is on The cover closed indicator may change repeatedly in course of tightening When it is nearly secured turn the handle slowly while observing whether the cover closed indicator is on o...

Страница 12: ...1 When the set sterilizing time is up the system will send out a sound alarm and the steam will be exhausted to release automatically EXHT indicator starts to blink 2 Under solid mode with drying when...

Страница 13: ...No sufficient water inside chamber makes heater dry scorch The elevation above 300m isn t set correctly on user administration menu Set correct elevation E04 Abnormal fluctuation of power results in a...

Страница 14: ...ill not be used for a long period of time the water in the sterilizer chamber must be emptied to avoid blocking the pipeline 3 Wipe the surface of the sealing ring and its contact parts with soft and...

Страница 15: ...tor 1 Under standby condition press and hold FUNC and DATA buttons then screen A will show 0000 Among these the first 0 flashing means it is changeable And press of UP or Down button may adjust the fi...

Страница 16: ...de P011 Version of system unmodified P012 Solid mode exhaustion temperature setting 105 C 130 C Chapter 8 Diagram of Pipeline Interface 1 A series diagram of pipeline 1 Pressure gauge 2 Pressure switc...

Страница 17: ...0A FD100A Electrical schematic diagram Drawell International Technology Limited Add Suite 1117 Lane561 XiuChuan Rd Greeland Max Mall PuDong Shanghai Tel 0086 21 54411195 Fax 0086 21 33823261 Web www d...

Отзывы: