User Manual of CIC-D100 Ion Chromatograph
56
1. Connect the pump outlet directly to the REGEN IN port of the suppressor. Connect the
REGEN OUT port directly to the waste liquid bottle. Inject deionized water at a flow rate of
0.3 mL/min for 10 min, and then seal with a plug;
2. Connect the pump outlet directly to the ELUENT IN port of the suppressor. Connect the
ELUENT OUT port directly to the waste liquid bottle. Inject deionized water at a flow rate
of 0.3 mL/min for 10 min, and then seal with a plug;
3. After sealing, leave for activation for not less than 30 minutes. Activation is completed.
Steps of replacing the suppressor:
1. Turn off the suppressor current and stop the pump;
2. Disconnect the suppressor cable and remove the pipe joint on the suppressor;
3. Remove the suppressor;
4. Install the new suppressor into the bracket (ELUENT OUT upward);
5. Connect the cables and reconnect all pipelines.
6. Start the pump and resume operation.
Figure 6-5 Replacement of the suppressor
:
The joint of suppressor cable should be connected properly, otherwise no
current will be applied.
6.9 Replacement of the power fuse
1. Turn off the main power switch of the instrument and unplug the "pin" power plug at the
Suppressor cable
Caution