Supplement Zeus
Infinity Empowered
Fitting the APL valve
1
Fit the APL valve (A) vertically on the upper part
of the breathing system.
The mark (B) must point in the direction of the
user when the breathing system is installed.
2
Tighten the knurled nut (C) clockwise.
Installing the flow sensors
1
Insert both flow sensors (A) in the breathing
system. Screw in place with the nozzles (only
use the blue, non-rotatable connecting
nozzles).
Fitting the flexible arm (optional)
1
If fitted: remove the bag elbow (A) for the
breathing bag from the breathing system.
2
Position the connector on the arm (B) on the
breathing system. Tighten it using the two
knurled screws. Check to make sure the arm is
secure.
006
156
WARNING
Risk of fire
Residual vapors of highly flammable disinfec-
tants (e.g., alcohols) and deposits that were
not removed during reprocessing may ignite
when the flow sensor is in use.
– Ensure particle-free cleaning and disinfec-
tion.
– After disinfection, allow the flow sensor to
air for at least 30 minutes.
– Before inserting the flow sensor check for
visible damage and soiling, such as resid-
ual mucus, medication aerosols, and parti-
cles.
– Replace flow sensors when damaged,
soiled, or not particlefree.
A
B
C
A
009
NOTE
Depending on the date of manufacture, it is possi-
ble that the breathing system does not have
threaded bushes for mounting the flexible breath-
ing bag holder. In this case, assembly is only pos-
sible if a new breathing system with threaded
bushings is used instead of the existing breathing
system.
A
B
C