22
Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of the calibration cradle).
Switch on the instrument and insert it into the calibration cradle.
Press the
[+]
key and keep it pressed for 5 seconds to open the calibration
menu, enter the password (password on delivery = 001).
Use the
[+]
key to select the single-gas adjustment function. The symbol
for span calibration
» «
flashes.
Press the
[OK]
key to start the channel selection.
–
The display shows the gas of the first measuring channel, e.g.
»CH
4
LEL«
.
Press the
[OK]
key to start the adjustment function of this measuring channel,
or use the
[+]
key to select another measuring channel (O
2
- Vol.-%, H
2
S - ppm,
CO - ppm, etc.).
–
The test gas concentration is displayed.
Press the
[OK]
key to confirm the test gas concentration or use the
[+]
key
to change the test gas concentration and complete the process by pressing
the
[OK]
key.
–
The measurement value flashes.
Open the test gas cylinder valve to let gas flow over the sensor with a
volume flow of 0.5 l/min.
–
The displayed, flashing measurement value changes to the value according
to the supplied test gas.
When the displayed measurement value is stable (after at least 120 seconds):
Press the
[OK]
key to perform the adjustment.
–
The display containing the current gas concentration changes with the
display
»OK«
.
Press the
[OK]
key or wait for approx. 5 seconds to end the adjustment of
this measuring channel.
–
The next measuring channel is displayed for adjustment if necessary.
–
After the adjustment of the last measuring channel, the instrument changes
to the measuring mode.
Close the test gas cylinder valve and remove the instrument from the
calibration cradle.
If a fault has occurred during the span calibration:
WARNING
Risk to health! Never inhale the test gas.
Observe the hazard warnings of the relevant Safety Data Sheets.
NOTICE
In the case of the CO
2
channel there is a 2-stage calibration routine:
The zero calibration takes place first and is then followed by the span calibration.
–
The fault message »
« appears and »
« is displayed for the
respective sensor instead of the measured value.
In this case, repeat the calibration.
Change the sensor if necessary.
Cleaning
The instrument does not need any special care.
Dirt and deposits can be removed from the instrument by washing it with
cold water. A sponge can be used for wiping if necessary.
Carefully dry the instrument with a cloth.
Maintenance
The instrument should be inspected and serviced once a year by suitably
qualified persons. Comparisons:
EN 60079-29-2 – Gas detectors - Selection, installation, use and
maintenance of detectors for flammable gases and oxygen
EN 45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for
selection, installation, use and maintenance
National regulations
Recommended calibration interval for measuring channels Ex, O
2
, H
2
S, SO
2
,
NO
2
and CO: 6 months.
Calibration intervals of other gases: see Instructions for Use of the respective
DrägerSensors.
See the Technical Manual for details of spare parts.
Test gas properties (such as relative humidity, concentration, etc.) can be found
on the corresponding data sheet of the sensor. The relative humidity is not
relevant for the O2 sensor. Always use dry test gas.
Storage
Dräger recommends storing the instrument in the charger module
(order no. 83 18 639).
Dräger recommends checking the charge of the power supply at least every
three weeks if the instrument is not stored in the charger module.
CAUTION
Abrasive cleaning tools (brushes etc.), cleaning agents and cleaning solvents
can destroy the dust and water filters.
Содержание X-am 5600
Страница 117: ...117 127 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger www draeger com ifu Dr gerSensor 90 33 890 X am 5600 X am 5600...
Страница 121: ...121 OK 3 3 2 1 ch4 UEG CAL 123 bt 2 A1 A2 TWA 1 STEL 1 H2S CO OK OK 3 2 1 1 Bump Test 5 21 O2...
Страница 122: ...122 A1 Dr ger IR Dr ger X am 5600 A1 O2 A1 A1 1 2 OK A2 OK STEL A2 O2 A1 A2 0 100 CH4 0 100 A2 STEL TWA...
Страница 146: ...146 EC LED Dr ger Dr ger X am 5600 A1 A1 A1 A1 LED A2 LED OK A2 OK A1 O2 A2 O2 A1 A2 0 100 Vol CH4 0 100 LEL...
Страница 151: ...151 OK OK OK 5 EN 60079 29 2 EN 45544 4 4 Ex O2 H2S SO2 NO2 CO 6 Dr gerSensor O2 Dr ger 83 18 639 Dr ger 3 Dr ger Dr ger...
Страница 179: ...179 OK OK 3 2 1 5 m 21 Vol 1 A1 Dr ger IR A1 O2...
Страница 180: ...180 A1 A1 A1 A2 OK A2 OK STEL TWA STEL TWA A2 O2 A1 A2 CH4 0 100 Vol 0 100 LEL A2 STEL TWA STEL 1...
Страница 181: ...181 OK 20 10 Dr ger OK 3 OK TWA STEL 10 2 1 3 PC Dr ger CC Vision 1 OK 60 1 CO2 XXS O3 PC Dr ger CC Vision N2...
Страница 183: ...183 0 5 l 83 18 752 IR Ex 20 1 IR CO2 20 1 O2 0 6 Vol 1 TOX 20 1 A1 A2 H2 PC H2 1 Dr ger 68 11 132 t90 90 33 890 90...
Страница 187: ......