27

Faults, Cause and Remedy

403

Measured value of the Dräger-

Sensor XXS H

2

S is in the 

negative range

Carry out the fresh air calibra-

tion, page 36.

404

DrägerSensor XXS H

2

S is not 

inserted

Check DrägerSensor XXS H

2

S, 

page 46.

405

Error during the function test 

with gas (bump test) of the 

DrägerSensor XXS H

2

S

Repeat the function test, cali-

brate or replace the 

DrägerSensor XXS H

2

S, if 

necessary, page 46.

407

Failed filter test.

Repeat filter test with X-dock.

408

Failed rise time test.

Repeat rise time test with X-dock.

501

The zero point calibration of 

the DrägerSensor XXS CO is 

not valid

Carry out the fresh air calibra-

tion, page 36.

502

The span calibration of the 

DrägerSensor XXS CO is not 

valid

Carry out the span calibration, 

page 40.

503

The measured value of the 

DrägerSensor XXS CO is in 

the negative range

Carry out the fresh air calibra-

tion, page 36.

504

The DrägerSensor XXS CO is 

not inserted

Check the DrägerSensor XXS 

CO, page 46.

505

Error during the function test 

with gas (bump test) of the 

DrägerSensor XXS CO

Repeat the function test, cali-

brate or replace the DrägerSen-

sor XXS CO, if necessary, 

page 46.

507

Failed filter test.

Repeat filter test with X-dock.

508

Failed rise time test.

Repeat rise time test with X-dock.

Special symbol 

»

«

 

and displayed numerical 

code:

Cause

Remedy

Содержание X-am 1700

Страница 1: ...ST 7452 2005 eps Multi Gas Monitor Technical Manual Dr ger X am 1700 2000 approved as type LQG 00xx...

Страница 2: ...the device 12 Before entering the workplace 13 During operation 13 Calling the Info Mode 14 Calling the Quick Menu 15 Possible functions of the quick menu 15 Calling the Calibration Menu 16 Calibratio...

Страница 3: ...n 33 Calibrating Adjusting the Device 35 Carrying out the fresh air calibration 36 Carry out the 1 button calibration 38 Calibrating adjusting the sensitivity for a single measuring channel 40 Replaci...

Страница 4: ...n hazard areas which have been tested and appro ved according to national European or international Explosion Protection Regulations may be used only under the conditions explicitly specified in the a...

Страница 5: ...measurement of four gases X am 1700 2 years service life from activation of the device refer to page 11 X am 2000 depending on the device type independent measurement of one gas up to four gases CAUTI...

Страница 6: ...s U so the year of construction is 2004 WARNING Read the safety measures in the Instructions for Use Do not replace or charge batteries in potentially explosive areas Danger of explosion D Type LQG 00...

Страница 7: ...o C depending on battery pack and batteries where gases or dusts of groups A B C D or E F G and temperature class T3 or T4 depending on battery pack and batte ries may be present Safety Instructions...

Страница 8: ...Alarm LED 3 Buzzer 4 key 5 key 6 Display 1 IR interface 2 Fastening clip 3 Type plate 4 Charging contacts 5 Power pack 1 Measured gas display 2 Measured value display 3 Special symbols 00223854_01 ep...

Страница 9: ...nd CO refer to page 14 The device is set to function test with gas bump test refer to page 29 The device is set to the fresh air calibration function refer to page 36 The device is set to the 1 button...

Страница 10: ...device on CD Dr gerSensor CatEx 125 PR LEL XXS O2 Vol XXS CO ppm XXS H2S LC ppm Measuring range 2 0 to 100 0 to 25 0 to 2000 0 to 100 Alarm A1 2 setpoint can be acknowledged self latching 20 yes no 19...

Страница 11: ...e Dr ger X am 2000 is ready for opera tion Only in the case of Dr ger X am 1700 The following activation sequence also has to be carried out Press and hold the key for approx 3 seconds until the count...

Страница 12: ...he calibration interval expires is displayed in days e g CAL 73 The sensor which is next due for calibration adjustment is displayed with the days remai ning until the next calibration adjustment e g...

Страница 13: ...nually switching the unit off and on again in fresh air After the measuring range of the TOX measuring channels has been exceeded temporarily up to one hour checking the measuring channels is not nece...

Страница 14: ...tivated Warning messages are displayed Numerical codes of the warning messages see page 23 key Fault messages are displayed Numerical codes of the fault messages see page 25 key The peak values the ma...

Страница 15: ...uick menu the device remains in measuring mode You can select the activated functions of the quick menu by pressing the key Press the key to call the selected function Possible functions of the quick...

Страница 16: ...ial symbol for the enter password function is displayed The display shows 000 with the first digit flashing Use the key to set the flashing digit Press the key the second digit starts flashing Use the...

Страница 17: ...s the key to cancel the active function If no key is pressed for 10 minutes the device returns automatically to measuring mode Fresh air calibration refer to page 36 1 button calibration refer to page...

Страница 18: ...ions oxygen channel or by manually switching the unit off and on again in fresh air STEL TWA exposure alarm The STEL and TWA alarm cannot be acknowledged or cancelled Switch off the device The values...

Страница 19: ...f necessary commission the Dr ger Safety Service Center to eliminate the error End of operation Only in the case of X am 1700 A warning period starts before the end of the operating time After switchi...

Страница 20: ...olutely necessary to flush the extension hose for a period of time to eliminate or minimize the effects which may interfere with measurements when using a sampling hose or a probe e g memory effects d...

Страница 21: ...tation and online help of the software A version of Dr ger CC Vision which can be used for X am 1700 2000 is supplied with the device on CD IR Calibration cradle order no 83 18 752 with inserted USB D...

Страница 22: ...GasVision is used for reading and displaying the data base Observe the documentation and online help of the software IR Calibration cradle order no 83 18 752 with inserted USB DIRA with USB cable orde...

Страница 23: ...repaired by Dr ger Service Special symbol and displayed numerical code Cause Remedy 151 Service life of device about to elapse Procure follow on device 152 Customer s service life coun ter about to e...

Страница 24: ...perature is too high Operate the device within the allowed temperature range 355 The temperature is too low Operate the device within the allowed temperature range 356 The calibration interval for Dr...

Страница 25: ...has elapsed Reset the service life counter using CC Vision 103 The device is defective The device must be repaired by Dr gerService 104 Check sum error program code The device must be repaired by Dr g...

Страница 26: ...zero point calibration of Dr gerSensor XXS O2 is not valid Carry out the fresh air calibra tion page 36 302 The span calibration of the Dr gerSensor XXS O2 is not valid Carry out the span calibration...

Страница 27: ...on of the Dr gerSensor XXS CO is not valid Carry out the fresh air calibra tion page 36 502 The span calibration of the Dr gerSensor XXS CO is not valid Carry out the span calibration page 40 503 The...

Страница 28: ...ered or after 2 weeks Device calibration page 35 In regular intervals according to the used sensors and the operating conditions For sen sor specific calibration data refer to the Instructions for Use...

Страница 29: ...over the sensors Wait until the device displays the test gas concentration with sufficient tolerance Ex 40 LEL 74 LEL 1 O2 17 4 vol 18 6 vol 1 H2S 12 ppm 18 ppm 1 CO 40 ppm 60 ppm 1 If the alarm setp...

Страница 30: ...st con centration Setting on delivery Quick bump test Prepare a test gas cylinder the volume flow must be 0 5 L min and the gas con centration must be higher than the alarm setpoint concentration to b...

Страница 31: ...OK The bump test that was carried out is do cumented with the result and date in the device memory Close the test gas cylinder valve and re move the device from the calibration cr adle If the concentr...

Страница 32: ...tration values have now fal len under the A1 alarm setpoints the de vice returns to the measuring mode If during the function test no alarm occurs within 1 minute the alarm mode is activa ted to indic...

Страница 33: ...ls participate in the function test Prepare the Bump Test Station according to the instructions Switch on the device and insert it into the receptacle of the Bump Test Station until it locks into plac...

Страница 34: ...s and the current measured values do not reach the set target concentration the er ror alarm is activated to indicate failure The fault message appears and is displayed instead of the measured value o...

Страница 35: ...ter 6 months Improve the zero point accuracy carry out the fresh air calibration page 36 Set the sensitivity of all sensors to the value of the test gas carry out the 1 button cali bration page 38 Se...

Страница 36: ...he message 109 is displayed with the special symbol for fault mes sage The message is cleared after 5 seconds and the function is available again in the menu During the fresh air calibration the zero...

Страница 37: ...he calibrati on or wait for 5 seconds If a fault has occurred during the fresh air calibration The fault message appears and is displayed for the respecti ve sensor instead of the measured value In th...

Страница 38: ...the test gas leaving the adapter into a fume cupboard or into the open air with a hose connected to the second connec tor of the calibration cradle Switch on the device and insert it into the calibra...

Страница 39: ...to the measuring mode Close the test gas cylinder valve and re move the device from the calibration cradle If a fault has occurred during the 1 button calibration The fault message appears and is dis...

Страница 40: ...alibration cradle Switch on the device and insert it into the calibration cradle Press the key and keep it depressed for 5 seconds to call the calibration menu enter the password and select the span c...

Страница 41: ...isplay OK Press the key to quit the calibration of this measuring channel The next measuring channel is offered for calibration After the last measuring channel has been calibrated adjusted the device...

Страница 42: ...ommended that a complete charging is carried out After batteries have been replaced The settings and data are stored when the battery is replaced The sensors warm up again WARNING Do not replace the b...

Страница 43: ...multiple charging station When attaching the charging modules disconnect the power pack from the mains supply Attaching charging modules 1 Turn the slots of the interlock into a horizontal position by...

Страница 44: ...it of the charging contacts in the charging modules e g by metallic objects that have fallen in does not result in damage to the charging station It should however be avoided due to possible heating h...

Страница 45: ...devices The power pack contained in the rechargeable battery and charging kit order no 83 18 785 is suitable for charging a device When using the vehicle charging adapter order no 83 17 754 it is reco...

Страница 46: ...laced from the sensor receptacle in the case of the Dr gerSensor CatEx 125 PR disconnect the flat cable connection before Insert the new sensor into the sensor re ceptacle Insert the flat cable connec...

Страница 47: ...tivity page 40 Electrochemical sensors Do not expose to fire Do not force open Danger Acid Burn Risk Like batteries only dispose of as special waste in accordance with local waste disposal regulations...

Страница 48: ...ts can be removed from the device by washing it with cold water A sponge can be used for wiping if necessary Carefully dab dry the device using a cloth NOTICE Abrasive cleaning implements brushes etc...

Страница 49: ...ugust 2005 and apply to this device Common household appliances can be disposed of using special collecting and recycling facilities However as this device has not been registered for household usage...

Страница 50: ...chargeable batteries Typically 12 hours under normal conditions Dimensions approx 130 mm x 48 mm x 44 mm H x W x D Weight approx 220 g to 250 g CE markings Electromagnetic compatibility Directive 2004...

Страница 51: ...s Measuring function for explosion protection and measurement of oxygen deficiency and surplus as well as toxic gases EXAM Essen Germany BVS 08 ATEX G 002 1 PFG 08 G 001 EN 60079 29 1 3 EN 50271 3 The...

Страница 52: ...ger X am 2000 Ex O2 H2S alkali 83 18 890 Dr ger X am 2000 Ex O2 CO H2S alkali 83 18 910 Dr ger X am 2000 CO H2S alkali 83 18 696 Dr ger X am 2000 O2 CO alkali 83 18 697 Dr ger X am 2000 O2 H2S alkali...

Страница 53: ...g module Dr ger X am 1 2 5000 max 2 83 15 635 Vehicle connecting line 12V 24V for Dr ger X am 1 2 5000 charging module 83 17 754 Vehicle installation set for 1 Dr ger X am 1 2 5000 charging module 83...

Страница 54: ...st Station including mixed gas cylinder 83 19 130 Dr ger Mobile Printer for the Bumptest Station 83 19 310 E Cal module Dr ger X am 1 2 5000 83 18 754 Calibration cradle Dr ger X am 1 2 5000 83 18 752...

Страница 55: ...55 Declaration of Conformity Declaration of Conformity...

Страница 56: ...00 en Dr ger Safety AG Co KGaA Edition 08 August 2012 Edition 01 February 2006 Subject to alteration Dr ger Safety AG Co KGaA Revalstra e 1 D 23560 L beck Germany Tel 49 451 8 82 0 Fax 49 451 8 82 20...

Отзывы: