Operational characteristics
Dräger VVP 1000
25
5.2
Touchscreen display
The buttons on the touchscreen display change dynamically
depending on the task being executed. To execute an action,
select the corresponding icon on the display.
5.3
Log user in or out
The following user levels are set up by default:
To log in a user:
1. Select .
a. Select .
b. Select the desired user level from the list.
2. Select the password input field.
3. Enter the password and confirm with .
To log out the current user:
1. Select .
Information about the current user will be displayed.
2. Select .
The current user will be logged out.
5.4
Viewing and acknowledging events
5.4.1
Viewing events
When an event occurs, this is indicated by the event symbol in
the header.
The following events are displayed and saved:
Maintenance urgently needed CHxxx
Maintenance necessary CHxxx
Exposure alarm CHxxx
Alarm 1 CHxxx
Alarm 2 CHxxx
Alarm 3 CHxxx
Alarm 4 CHxxx
Average alarm 1 CHxxx
Average alarm 2 CHxxx
Average alarm 3 CHxxx
Average alarm 4 CHxxx
Interruption due to PC CHxxx
Sensor warming up CHxxx
Over range CHxxx
Under range CHxxx
To display events:
Select
in the header or select >
Events
The Events screen shows the events in different colours:
5.4.2
Acknowledging events
To acknowledge an event:
Select the desired event from the Events list.
The acknowledged message turns blue.
To acknowledge all events of the Events list and the control unit:
Select .
User
Password
Configurator
0001
Administrator
0002
Service
XXXX
NOTICE
If the touchscreen is not operated for 10 minutes, the
current user will be logged out automatically.
i
i
00233547.eps
1
Colour
Meaning
Red
Gas alarm
Yellow
Fault or warning
Black
Status information
Blue
Acknowledged message
00333547_en.eps