2.2
Intended Use
The Dräger VVP 1000 (hereafter referred to as "display") is a
display for the graphical visualisation of up to 100 channels of
a VarioGard system.
2.3
Limitations on use
The display is not designed or authorised for use in areas
where flammable or explosive gas mixtures could develop.
3
Approvals
Safety
IEC 61010-1, EN 61010-1: Safety requirements for electrical
components, Part 1.
IP66 degree of protection classification (front only), IEC 529
Type 4X degree of protection classification (front only), UL50
Electromagnetic compatibility
Emissions and immunity for EN 61326: Electrical equipment
for measurement, control and laboratory use.
Immunity to industrial locations:
RF interference
EN 61000-4-6 Criterion B
3 V/rms
Surge voltage
EN 61000-4-5 Criterion A
1 kV L-L,
2 kV L&N-E power
Fast transients (burst) EN 61000-4-4 Criterion A
2 kV power
1 kV signal
Power frequency
magnetic fields
EN 61000-4-3 Criterion A
10 V/m
Electrostatic
discharge
EN 61000-4-2 Criterion A
8 kV air discharge
4 kV contact discharge
Emissions:
EN 55011
Class A
FCC conditions
This device has been tested and complies with Part 15,
Class B of the FCC conditions. Its operation is subject
to the following conditions:
(1) This device must not cause harmful interference.
(2) This device must accept any received interference;
this includes interference caused by unintentional
operation.