Primus Workbook

Workbook
V2.1

Primus 

W

or

kbook 

/ V2.1 

/ MR00092 

/ 25.01.06 

/ Printed 

in 

Ge

rm

an

/ Chlorine-free 

en

vironmentally 

compa

tible 

/ Subject 

to 

modifica

tions 

/ © 

2006 

Dräger 

Medical 

A

G

 & 

Co

.K

G

The quality management system at Dräger
Medical AG & Co. KG Education and
Training complies with ISO 9001:2000

Draeger Medical UK Ltd

The Willows, Mark Road
Hemel Hempstead
Herts, HP2 7BW

T:  01442 213542
F:  01442 240327
E:  [email protected]
W: www.draeger.co.uk

Primus Family Quick Guide

Содержание Primus Series

Страница 1: ...ormance characteristics of the Dräger Medical devices described in this Quick Guide each user must first carefully read and fully comprehend the Instructions for Use before operating the device Primus Workbook Workbook V2 1 Primus Family Quick Guide ...

Страница 2: ...2 A quick guide to provide information on the basic functions of the Primus This guide presupposes a knowledge of the Primus Instructions for Use and is not intended to replace them ...

Страница 3: ...tion 10 Monitoring 10 Setting Fresh Gas Concentrations 10 Selecting a Ventilation Mode 12 Spontaneous Breathing SPONT 12 Manual Ventilation MAN 12 Mechanical Ventilation 13 Switching to Standby 15 End of Operation 15 Switch Off 15 Emergency Scenarios 16 Emergency Start Ups 16 Power Failure 17 Gas Supply Failure 18 Ventilator Failure 19 Mixer Failure 20 Cleaning 21 Consumable Items 24 Contents ...

Страница 4: ...ral supply wall outlets Observe LEDs for the supply pressure they must be green 2 Open back up gas cylinder valves to check cylinder pressure The LEDs for the O2 Air and N2 O cylinder pressure must be green 3 Close the cylinder pressure valves again Checking Electrical Supply Ensure that main power cord is plugged in Green LED is illuminated whenever main power is available 4 4 1 2 3 4 Preparation...

Страница 5: ... bag and patient hoses In Addition Test the oxygen flush 3 Check oxygen safety control 4 Make sure vaporisers are in position locked and filled 5 Check water trap level and the sample gas line connection 6 Check function of suction system 7 Your suction system may differ from the system shown Connect anaesthetic gas scavenging system 5 1 34 2 6 7 5 Preparation for Use ...

Страница 6: ...the APL valve Your version may differ from the one shown Close oxygen safety control valve 3 Confirm the checklist if all points are OK Primus will run internal system tests Note During this start up procedure and up until the checklist is confirmed manual ventilation is always possible The oxygen safety flow can be set to 0 12 L min and will be channelled through the vaporiser 6 2 3 1 Start Up ...

Страница 7: ...Faulty breathing hoses or breathing bag Sample gas line Filter Incorrectly connected absorber Flow sensors Breathing system Vaporisers Water trap Oxygen sensor Potential Test Results All test results are colour coded and are displayed next to each specific test function An advisory window containing information on how to remedy the problem accompanies any error detected during the self test 7 1 St...

Страница 8: ...splayed see page 7 Self Test Colour Result Action Example Outcome Green Test completed No action 100 functional successfully required Yellow Primus can be Faults must be Loudspeaker failure used with reported or remedied restrictions immediately after use Red The device has If the fault is recognised O2 failure a malfunction immediately and can be instantly corrected the test is repeated If operat...

Страница 9: ...fault settings Set alarm limits Self test results display Perform the leak test Delete trends Default configuration password protected held by Key Users and Dräger Access monitoring mode Monitoring Mode In monitoring mode the Primus will measure and display gas measurements There is no fresh gas delivery or ventilation capability If ventilation is required set the appropriate gas flow select the a...

Страница 10: ...entrations These may be preset in standby Carrier gas press and confirm either N2 O or Air 1 Oxygen concentration press O2 set the concentration with the rotary knob and confirm 2 Fresh gas flow press flow l min set desired flow and confirm 3 Fresh gas flow will start as soon as a ventilation mode is activated and is indicated in virtual flow tubes on display screen 10 3 1 2 1 Operation ...

Страница 11: ...mus is fitted with a hypoxic guard function SORC This is an electronic O2 minimum delivery system that guarantees at least 25 O2 whenever N2 0 is used as the carrier gas This prevents the delivery of hypoxic gas mixtures 11 Operation ...

Страница 12: ... knob Manual Ventilation MAN Switch APL pressure limiting valve to MAN There are two types of APL Please familiarise yourself with your own device Press Man Spont key and confirm with the rotary knob Note the breathing system can be quickly depressurised by pressing the lever of the APL valve or pulling the APL valve upwards dependant on the type of APL valve installed 12 Operation ...

Страница 13: ...rd key Press rotary control knob to confirm Fresh Gas Outlet optional The Primus has an external fresh gas outlet to allow the user to use non rebreathing systems such as a Bain circuit or Ayres T piece Prepare the circuit in accordance with the manufacturers instructions Attach the circuit to the external fresh gas outlet located on the left hand side of the Primus Press the external fresh gas ou...

Страница 14: ...d Press the alarm limits soft key Use the rotary control to select the alarm limit you wish to change Confirm change by pressing the rotary control Auto Set Alarm Limits The auto set limits soft key can be used to adapt minute volume and airway pressure when a ventilation mode is selected Operation ...

Страница 15: ...press hard key and press rotary knob to confirm To switch off Primus press the ON OFF knob You will hear a series of beeps when you switch the Primus off During these beeps you can switch the Primus back on and the device will continue in the mode it was in Operation ...

Страница 16: ...ation is available at any time Switch on Primus After approximately 35 seconds the checklist will appear Press the cancel test key 1 You can cancel the self test up to 10 times After that it is mandatory to perform a self test Primus switches to standby mode after completing the minimal test duration 10 seconds Close safety O2 flow control Continue as in normal operation 1 Emergency Scenarios ...

Страница 17: ...ttery capacity will be shown on the screen When the battery is almost empty the message BATTERY LOW will be displayed If the battery is completely empty the device automatically switches to Man Spont Manual ventilation is available at any time Release the oxygen safety control knob by pressing and set to desired oxygen flow between 0 and 12 l min The oxygen flow is channelled through the vaporiser...

Страница 18: ...he gas mixture previously delivered the amount of fresh gas flow l min will remain the same 100 Air will be delivered if O2 fails 100 O2 will be delivered if N2 O or air fails Open the appropriate backup gas cylinder valve for O2 or N2 O or reconnect the central gas supply If the carrier gas N2 O fails you are able to use Air and vice versa Emergency Scenarios ...

Страница 19: ...the ventilator will stop and display the error message VENTILATOR FAIL on the screen Fresh gas flow and full monitoring functionality are not affected Operation automatically switches to Man Spont only manual ventilation is possible Emergency Scenarios ...

Страница 20: ...on the screen Ventilation and full monitoring functionality are not affected Release the oxygen safety control knob by pressing and set to desired oxygen flow between 0 and 12 L min The oxygen flow is channelled through the vaporiser Emergency Scenarios ...

Страница 21: ...ick guide to disassembling reassembling the breathing system Disassembly instructions Removing the Breathing System from the Primus Remove accessory items such as flow sensors breathing circuit filters etc Remove the inspiratory and expiratory ports Remove the Breathing System from the Primus as shown below Cleaning 21 ...

Страница 22: ...g the Allen Key B Diagram 1 Remove the cover Lift off the metal heater plate Loosen the 3 remaining sealing screws to access the ventilator diaphragm and remove Diagram 2 Prepare all parts for cleaning disinfection process as per Hospital policy Diagram 1 Diagram 2 Cleaning ...

Страница 23: ...or diaphragm Replace the breathing block A and tighten the 3 sealing screws Replace the metal heater plate B Replace the cover and tighten the 5 sealing screws C Replace the inspiratory and expiratory ports Replace the accessory items Cleaning 23 ...

Страница 24: ... 150cm Latex Free MP01805 Twinstar HME Filter 50 MP01850 Ergostar CM50 catheter mount 50 8290286 Sample Line 10 8403735 Spirolog flow sensor 5 8401892 Adult Test Lung 1 MX00001 Drägersorb 800 Soda 1 Lime 5 litre MX00004 Drägersorb Soda Lime Clic 6 Canister 1 3 litre 6872130 Watertrap 12 MP00301 8403735 8401892 MX00001 MX00004 6872130 MP01805 MP01850 8290286 ...

Страница 25: ...25 Notes ...

Страница 26: ...26 Notes ...