Quick Reference Guide Primus
Infinity Empowered
SW 4.5n
17
Getting Started
Getting Started
Device check
Power on
1
Switch on Primus
Infinity Empowered
: press the
system power switch
(A), an acoustic tone
sounds.
After approx. 15 seconds all LEDs and the loud-
speaker are tested by Primus
Infinity Empowered
.
The initial screen appears after about 20 seconds. Pri-
mus
Infinity Empowered
now loads its software and
tests its internal memory.
Check List
The
Check List
is displayed after about 35 seconds.
The test steps to be performed are grouped in five cat-
egories:
A
Gas Supply
– pipeline pressure
– cylinder pressure
– O
2
flush functional?
– Safety O
2
control functional?
B
Vaporizers
– Correctly locked in positon?
– Set to zero?
– Fill level OK?
– Safety filler locked?
C
Breathing Circuit
– Fully assembled?
– Correctly connected?
– Gas scavenger connected and flow adjust-
ed?
– Soda lime OK?
D
Miscellaneous
– Water trap fill level OK?
– Suction OK?
– Emergency resuscitator present and func-
tional?
NOTE
The Primus
Infinity Empowered
Quick Reference Guide is not a replacement or substitute for the Instructions for
Use. Any use of the device requires full understanding and strict observation of the Instructions for Use.