Assembly
Assembly
Before first use
Charging the battery for emergency
operation
Primus
Infinity Empowered
has a built-in uninterruptible
power supply UPS which maintains the power supply for
at least 30 minutes (up to 90 minutes, depending on the
ventilation parameters) in the event of a mains power
failure, provided that the battery is charged.
Switching to battery power (UPS) takes place automat-
ically, and is indicated on the screen by the message:
POWER FAIL.
The battery recharges automatically when the worksta-
tion is plugged into the mains, but only up to a maximum
ambient temperature of 35 °C (95 °F).
The battery must be charged for 10 hours before using
the workstation for the first time:
z
Plug the mains power plug of the Primus
Infinity Empowered
workstation into the mains sock-
et.
The mains voltage must correspond to that specified on
the rating plate on the back of the machine.
The green LED
(A) lights up.
Leave Primus
Infinity Empowered
connected to the
electrical power supply for 10 hours. It does not need to
be switched on.
Connecting the gas supply
1
Screw on the compressed gas hoses of the central
gas supply (CS) for
O
2
,
AIR
,
N
2
O
to the front con-
nections of the gas inlet block (A). The two outer
ports at the back are reserved for the reserve gas
cylinders.
A compressed air outlet for the optional secretion
aspiration and an O
2
outlet for an external O
2
flow
tube are optionally available.
2
Plug the other end of the pressure hoses into the ter-
minal unit.
3
Make sure all supplies are connected correctly and
functioning properly.
NOTE
The Primus
Infinity Empowered
Quick Reference Guide is not a replacement or substitute for the Instructions for
Use. Any use of the device requires full understanding and strict observation of the Instructions for Use.