Maintenance
76
Instructions for use
Polaris 600
Preventive maintenance
Table of preventive maintenance intervals
Repair
It is recommended that only original parts from
Dräger are used and that the parts are replaced by
Dräger.
WARNING
Risk of faulty components
Device failure is possible due to wear or
material fatigue of the components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
WARNING
Risk of electric shock
Disconnect all electrical connections from the
power supplies before carrying out any work.
Component
Interval
Task
Personnel responsible
Locking segment of the
spring arms
Every 2 years
Check the dimensions of
the locking segment
(minimum material thick-
ness and minimum
length) and replace if
necessary.
Specialized service per-
sonnel