Operating concept
Correct operation of the sterilizable handle E
The touch-sensitive part of the handle is designed
so that, for correct operation, only one finger must
touch the operating section at any one time. If
several fingers touch the operating section
simultaneously, the operating signal will not be
implemented.
Correct operation
– One finger at a time is touching the operating
section
Operating signal will not be implemented
– Several fingers are touching the operating
section at the same time
Wall-mounted control panel (option)
The wall-mounted control panel can incorporate up
to 3 light control panels. Optionally it can include a
camera control panel.
The keys on the wall-mounted control panel are
pressure-sensitive, i.e., a certain pressure must be
applied by the finger.
26658
26659