Initial situation and consequences
Initial situation according to the customer letter “COVID-19: Usage of Dräger anesthesia devices for long-
term ventilation”. That means:
high fresh gas flow 1.5xMV
no volatile anesthetic agent
soda lime at the device
active AGS
manual breathing bag connected
mechanical filters with HME connected
Potential impact:
High gas supply consumption according to the set fresh gas flow (150% of the minute volume)
Compressed air consumption due to the active AGS additional to the fresh gas flow
IV medication needed for sedation
Low soda lime consumption
Drying out of the soda lime, but no patient harm (e.g. compound formation), as no volatile
anesthetic agent must be used
No room air contamination
Reduced humidity in the circle system
2
Shortage in IV medication for sedation
Use alternative IV drugs for sedation
Avoid usage of volatile anesthetic agents
3
Central gas supply shortage
Reduce the fresh gas flow, see chapter 12
If possible, connect gas cylinders to the device
If one gas (O
2
or AIR) fails, fresh gas supply can continue with the remaining gas. However, it
is not possiblbe to adjust the fresh gas O
2
concentration any longer.