NM2B
PMC PROCEDURE (continued)
6.31 FINAL TESTS
( )
6.31.1
Operator’s Instruction Manual
6.31.1.1
Verify that the availability/location of the machine’s Operator’s
Instruction Manual is in close proximity of the machine.
( )
6.31.2
Auxillary Lamp - if applicable
6.31.2.1
Verify that the lamp head (shroud) can move approximately 90°.
6.31.2.2
Verify that the lamp head (shroud) can be positioned at any point
within the 90° range and holds fast.
6.31.2.3
Verify that the lamp can rotate 360° and hits a stop located within
45° on either side of the top center point.
6.31.2.4
Verify that the lamp is plugged into a live AC receptacle.
6.31.2.5
Turn the lamp ON.
6.31.2.6
Adjust the dimmer control and verify that the light varies with the
adjustment of the knob.
6.31.2.7
Turn the lamp OFF.
( )
6.31.3
Final Check
6.31.3.1
Verify that all cylinder pressure gauges indicate zero.
6.31.3.2
Verify that the pipeline hoses are connected to the hospital
pipeline.
6.31.3.3
Verify that the APL valve knob is turned completely counter-
clockwise (fully open).
6.31.3.4
Place the Auto/Man selector in the BAG position.
6.31.3.5
Verify that the O
2
Med sensor is removed from the valve dome
adapter.
6.31.3.6
Verify that the valve dome is plugged.
6.31.3.7
Verify that the machine is plugged into a live outlet.
6.31.3.8
Return all machine controls and settings to their original state.
Rev. Y
6-61
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...