PMC PROCEDURE (continued)
NM2B
6.30 OXYGEN FLUSH VALVE
Turn the System Power switch to STANDBY.
6.30.1
Press and release the O
2
FLUSH button.
6.30.2
Does the flow of oxygen stop immediately? __ (Y)
6.30.3
Close the APL valve.
6.30.4
Connect a 12-inch hose to the inspiratory valve.
6.30.5
Set the Man/Auto selector to BAG.
6.30.6
Occlude the bag mount.
6.30.7
Insert the sensor from a calibrated O
2
Med into the valve dome adapter
on the inspiratory valve.
6.30.8
Close all flow control valves.
6.30.9
Press the O
2
FLUSH button.
( )
6.30.10
What is the O
2
concentration after the value stabilizes?___%O
2
(97-100)
6.30.11
Remove the O
2
Med sensor and install the plug.
6.30.12
Remove the test minute volumeter (P/N 2212300) from the absorber and
connect it to the common gas outlet, using Fresh Gas Outlet Volume
Test Device (P/N S010158).
6.30.13
Press and hold the O
2
FLUSH button for 15 seconds; multiply the value
by 4.
( )
6.30.14
What is the oxygen flush flow rate? ___l/min. (45-65)
6.30.15
Remove the test minute volumeter and test fixture, and reconnect the
fresh gas hose.
6.30.16
Turn the System Power switch to ON.
Rev. Y
6-60
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...