PMC PROCEDURE (continued)
NM2B
6.25B
BELLOWS: PEDIATRIC INTERNAL - if applicable
6.25B.1
Set the ventilator frequency to 10 BPM.
6.25B.2
Set the I:E Ration to 1:2.
6.25B.3
Turn the ventilator ON.
6.25B.4
Adjust the Inspiratory Flow control on the ventilator to Medium.
6.25B.5
Set the O
2
to 3 l/min.
6.25B.6
Occlude the 15mm patient side of the Y-piece and fill the bellows by
pressing the O
2
Flush button.
6.25B.7
Observe the absorber breathing pressure gauge as the ventilator cycles.
6.25B.8
What is the pressure when the bellows completes its downward travel?
___cmH
2
O >30
6.25B.9
What is the pressure when the bellows completes its upward travel?
___cmH
2
O
≤
3
6.25B.10 Install a 3 L breathing bag onto the patient circuit Y-piece.
6.25B.11 Fill the bellows by pressing the O
2
Flush button.
6.25B.12 Adjust the O
2
flow to 10 l/min.
( )
6.25B.13 What is the PEEP? ___cmH
2
O (
≤
3)
6.25B.14 Adjust the O
2
flow to 500 ml/min.
6.25B.15 Adjust the inspiratory flow control until the bellows collapses entirely.
( )
6.25B.16 Does the ventilator deliver the full tidal volume during 5 consecutive
cycles? ___(Y)
6.25B.17 Set the ventilator frequency to 20 BPM.
6.25B.18 Adjust the O
2
flow to 5 l/min.
6.25B.19 Adjust the inspiratory flow control until the bellows collapses to the 100
ml mark on the pediatric bellows assembly.
Rev. Y
6-56
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...