PMC PROCEDURE (continued)
6.7D
SUCTION REGULATOR
6.7D.1
Verify that the suction bottle is attached to the suction regulator.
6.7D.2
Verify that vacuum is attached to the ¾ in. DISS vacuum connection.
6.7D.3
Set the vacuum on/off valve to the OFF (vertical) position.
6.7D.4
Connect a digital pressure meter to the collecting inlet stem of the
suction bottle.
6.7D.5
Set the digital pressure meter to the mmHg scale.
6.7D.6
Turn the vacuum control knob fully counter-clockwise.
( )
6.7D.7
What is the vacuum indicated on the digital pressure meter? (0)
6.7D.8
Turn the vacuum control knob fully clockwise and verify that the
vacuum control knob stops.
6.7D.9
Set the vacuum on/off valve to the ON position.
6.7D.10
Set the regulator to indicate 250 mmHg.
( )
6.7D.11
What is the vacuum indicated on the digital pressure meter? (200-300
mmHg)
6.7D.12
Return all controls to their original positions.
Rev. Y
6-13