NM2B
REPLACEMENT PROCEDURES (continued)
4.30
Oxygen Sensor
The oxygen sensor is located on top of the inspiratory valve. Figure 4-30 shows
the arrangement of the sensor capsule and its housing, and also its connection
to the interface panel.
4.30.1
Set the System Power switch to
STANDBY
.
4.30.2
Pull the oxygen sensor housing from the inspiratory valve dome. (It
is a press fit.)
4.30.3
Unscrew the cover from the sensor housing and remove the sensor
capsule.
4.30.4
Remove the replacement sensor capsule from its shipping container
and install it in the housing. Ensure that the copper rings on the
capsule mate with the electrical contacts in the sensor housing.
4.30.5
Wait 15 minutes to allow the sensor capsule to stabilize.
4.30.6
Restore power to the machine and perform the 21% calibration
procedure for the oxygen sensor given in Section 5.
4.30.7
Press the sensor assembly into the inspiratory valve dome.
4.30.8
Perform the PMS Procedure given in Section 6.