80
Routine Maintenance
Maintenance of the UV Lamp
Removing and Replacing the UV Lamp
To remove the UV lamp from your instrument:
1. Ensure the instrument is turned off.
2. Remove the lamp housing cover.
3. Tilt the Multi-PID 2 slightly and remove the UV lamp.
4. Without touching the lamp window, place the new lamp into the
Multi-PID 2 lamp holder, window first. See Figure 19.
WARNING:
Do not remove or replace the UV lamp in a hazar-
dous location, danger of an explosion!
WARNING:
You must turn the instrument off before removing
the lamp cover.
WARNING:
Do not touch the wire grid inside the detector cell.
Any dust or dirt in the detector cell can be blown out
with a gentle jet of compressed air.
Do not insert any object, other than the UV lamp,
into the lamp holder.