Calibrating the instrument
1. In measuring mode, press the
key three times.
If functions in the quick menu have been activated using the
Dräger CC-Vision PC software, you can select these functions
using the
key. If no functions have been activated in the
quick menu, the instrument remains in measuring mode.
2. You can select the activated functions of the quick menu by
pressing the
key.
Press the
key to call the selected function.
Press the
key to cancel the active function and to switch to
measuring mode.
If no key is pressed for 60 seconds, the instrument reverts
automatically to measuring mode.
5
Calibrating the instrument
Adjustment may not be possible due to instrument and channel
errors.
5.1
Performing a fresh air calibration
Calibrate the instrument to fresh air, free of measured gases or
other interfering gases.
The zero point of the sensor is set to 0 during the fresh air
calibration.
1. Switch on the instrument.
2. Press the
key 3 times; the symbol for fresh air calibration
» «
appears.
3. Press the
key to start the fresh air calibration.
The measured values flash.
When the measured values have stabilized:
4. Press the
key to carry out the calibration.
The display containing the current gas concentration changes with
the display
» OK «
.
5. Press the
key to exit the calibration or wait for approx.
5 seconds.
If a fault has occurred during the fresh air calibration:
The fault message
» «
appears and
»
«
is displayed for the
respective sensor instead of the measured value.
Repeat the fresh air calibration.
If necessary replace the sensor; see the Technical Handbook.
5.2
Performing a standard sensitivity calibration
During the sensitivity calibration, the sensitivity of the sensor is set to
the value of the test gas.
1. Fit the calibration adapter (68 06 291) to the sensor cap.
2. Connect the test gas source to the calibration adapter.
3. Vent the test gas into a fume cupboard or into the open air
(with a hose connected to the second connector of the calibration
adapter).
4. Switch on the instrument.
5. Press the
key and keep it pressed for 5 seconds to call the
calibration menu.
WARNING
Always adjust the zero point first, before the sensitivity,
otherwise the adjustment will be incorrect!
OK
!
OK
CAUTION
Never inhale the test gas. Danger to health!
Observe the hazard warnings of the relevant Safety Data
Sheets.
NOTICE
To minimise adsorption effects, keep the length of the hose
as short as possible (maximum hose length: 1 m). Dräger
recommends using PTFE hoses.
OK
OK
!
i
i
Содержание MQG 0020
Страница 2: ......
Страница 28: ...28 Dr ger X am 5100 Technical Data...
Страница 69: ...Dr ger X am 5100 69 Dados t cnicos...
Страница 158: ...158 Dr ger X am 5100 1 6 170 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Страница 163: ...Dr ger X am 5100 163 3 1 5 1 2 Bump Test 21 O2...
Страница 166: ...166 Dr ger X am 5100 166 3 6 3 6 1 A1 A1 A1 1 2 3 6 2 A2 A1 i i OK...
Страница 169: ...Dr ger X am 5100 169 5 2 1 68 06 291 2 3 4 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 1 Dr ger PTFE i i OK OK OK OK OK OK...
Страница 170: ...170 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Страница 220: ...220 Dr ger X am 5100 1 6 232 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Страница 225: ...Dr ger X am 5100 225 3 1 5 1 2 Bump Test 21 O2...
Страница 227: ...Dr ger X am 5100 227 A1 A2 8 A1 230 3 4 A1 Dr ger X am 5100 1 0 5 l min 2 68 06 291 3 4 5 6 230 7 8 A1 OK OK...
Страница 228: ...228 Dr ger X am 5100 9 A1 3 5 228 3 6 3 6 1 A1 A1 A1 A1 A2 3 6 2 A2 A1 i i OK...
Страница 230: ...230 Dr ger X am 5100 4 2 4 3 1 Dr ger CC Vision 2 60 5 5 1 0 1 2 3 3 4 OK 5 5 Bump Test Bump Test 226 230 OK OK OK OK...
Страница 231: ...Dr ger X am 5100 231 5 2 1 68 06 291 2 3 4 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 1 m Dr ger PTFE i i OK OK OK OK OK OK...
Страница 232: ...232 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr ger 6 2 7 Dr ger Dr ger...
Страница 259: ...Dr ger X am 5100 259 1 6 271 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Страница 264: ...264 Dr ger X am 5100 3 1 4 3 2 1 3 1 5 1 2 Bump Test Bump Test OK 21 Vol O2...
Страница 270: ...270 Dr ger X am 5100 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 OK OK OK OK OK OK...
Страница 271: ...Dr ger X am 5100 271 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Страница 286: ...286 Dr ger X am 5100 Notes on Approval A 2 Declaration of Conformity...
Страница 287: ......