Overview
The Infinity Trident (NMT)
pod is a stimulator pod for the measurement of
neuromuscular transmission (NMT) / neuromuscular blockade in adult or pediatric
mode.
The NMT pod provides automatic measurements of muscle response (thumb twitch)
to electrical stimuli transmitted via electrodes placed over the peripheral nerve (for
example Ulnar nerve). The NMT sensors measure muscle response as well as skin
temperature and send this information to the pod and monitor for display. An
accelerometer is used as the twitch sensor. A thermistor is used to measure skin
temperature. It is important to understand the pulse characteristics of the stimulator
(see monitoring modes on page 24-5 and specifications in appendix B for details).
At the start of NMT measurements, the pod sends several test pulses to the nerve in
order to establish a supramaximal current, which is the pulse strength necessary to
excite all the fibers in the nerve. Once the supramaximal current has been established,
the pod performs a reference measurement consisting of four pulses to establish a
reference muscle response level (reference twitch) in the unrelaxed patient. In
subsequent measurements, the pod can compare the relaxed patient’s muscle
responses to this reference twitch.
NMT values are displayed in the NMT parameter box and saved as trend values. You
can set NMT alarm limits on the Alarm Limits table (see page 24-5).
NOTE:
The NMT pod is not supported in neonatal mode.
Only EEG and BISx pods can be connected to the NMT pod.
NOTE:
The supramaximal current and reference twitch should be established before
you administer muscle relaxants to the patient.
If the pod cannot establish a supramaximal current based on an actual
measurement, it uses a set reference value of 60 mA.
If the pod cannot establish a reference twitch, it uses an internal reference value
for twitch response. A reference twitch is necessary for single-mode
measurements.
Patients with nerve damage or other neuromuscular problems may not respond
properly to stimulation. The NMT pod may show unusual patterns when
monitoring relaxation in these patients.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...