![Dräger Infinity Delta Скачать руководство пользователя страница 342](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779342.webp)
19-12
D
ELTA
/D
ELTA
XL/K
APPA
VF8
19
ET
CO2 (E
ND
-T
IDAL
CO2)
MONITORING
Quick Reference Table – etCO
2
Setup
The following table explains etCO
2
setup functions.
Function
Description
Settings
Calibration
Sensor Cal.
Displays date and time of last Capnostat sensor
calibration
Not applicable
Adapter Cal.
Initiates airway adapter calibration
Not applicable
Gas Compensation
Gas compensation offsets inappropriate levels of anesthetic agents by ensuring that gas
percentages in each respiratory phase add up to 100: each gas in any gas mixture exerts a partial
pressure of the total. Room air, for example, is made up of approximately 79% nitrogen and 21%
oxygen. Gas concentration is usually expressed as a percentage, while partial pressure is
measured in mmHg or kPa. These setup functions differ according to the device used to monitor
etCO
2
(pod or module).
Agent
Compensation
(pod only)
Sets value of primary expired anesthetic agent
• 1 % - 20 %
Gas
Compensation
(pod only)
Determines gas percentages for inspiratory and
expiratory breathing. The table below lists
settings on the Gas Compensation menu in the
first column and lists automatically calculated
percentages in the right five columns
• Air
• N2O/O2
• >60% O2
• Heliox
N2O
Compensation
(module only)
Compensates for presence of nitrous oxide, not
normally present in room air but typically used in
the operating room. Significant concentrations
of nitrous oxide can cause the monitor to
overestimate the level of etCO
2
by
approximately 5 %.
• ON
• OFF
O2 Compensation
(module only)
Compensates for patient’s supplemental
oxygen. Failure to compensate for supplemental
oxygen can also result in inaccurate
measurements.
• 21 % - 100 %
Other etCO
2
setup functions
Scale
Determines size of currently displayed
waveform
• 40 mmHg
• 60 mmHg
• 80 mmHg
•
100 mmHg
(pod only)
Setting
Agent
CO2 O2 N2
N2O He
Air
5
17
78
0
0
N2O/O2
5
35
0
60
0
>60% O2 5
75
20
0
0
Heliox
5
35
0
0
60
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...