Status Messages
Message
Possible Cause
Suggested Action
tpO
2
> #
tpO
2
< #
tpO
2
or tpCO
2
values are outside the
alarm limits, due to one of the following
conditions:
• Physiological
• Inappropriate limit settings
• Loose fixation ring
• Check the patient and treat, if
necessary.
• Change the alarm limits (see
Alarms
chapter).
• Apply a new fixation ring at different
monitoring site.
• Check the sensor and pod;
recalibrate the sensor; change the
sensor membranes or replace the
sensor, if necessary.
tpCO
2
> #
tpCO
2
< #
TP Out of Range
(High)
tpO
2
/CO
2
values are outside the
measuring range, due to one of the
following conditions:
• Physiological
• Equipment malfunction
• Check the patient and treat, if
necessary.
• Check the sensor and pod;
recalibrate the sensor; apply new
sensor membranes; replace the
sensor, if necessary.
TP Warming Up
-- Reduced
Accuracy.
Warm-up phase in progress
Wait until the warm-up phase has
ended before relying on the displayed
tpO
2
/CO
2
values for clinical
evaluation.
TP Sensor
Unplugged
The tpO
2
/CO
2
sensor is unplugged.
• Check all connections.
• Replace the sensor and recalibrate,
if necessary.
TP Pod
Unplugged
The tpO
2
/CO
2
pod is not connected to
the monitor.
• Check all connections.
• Replace the pod or connecting cable,
if necessary.
Check TP
Sensor Site
The site timer will expire within 15
minutes.
Prepare to recalibrate and reposition
the sensor.
TP Check/
Replace Sensor
The auto heater shutdown is in effect or
the sensor didn’t heat to the desired
temperature setting.
• Recalibrate sensor and apply at new
monitoring site to start the Site
Timer.
• Verify that the sensor is compatible
with pod.
TP Change
Sensor Site
The site timer has expired.
• Immediately remove the sensor from
the patient.
• Recalibrate the sensor and apply at
new monitoring site.
Remove Probe
from Chamber
The calibration of the sensor is
complete.
Remove the sensor from the
calibration chamber.
TP Chamber
Timeout
The electrode remained in the
calibration chamber > 30 min.
Recalibrate or cover the sensor with a
protective cap for storage.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...