Status Messages
Message
Possible Cause
Suggested Action
<xx> S <#>
<xx> D <#>
<xx> M <#>
The pressure value is outside the
alarm limits.
• Check the patient and treat, if necessary.
• Access the Alarm Limits menu and change
the alarm limits.
• Check the equipment and replace it, if
necessary.
<xx> Out of
Range (High)
<xx> Out of
Range (Low)
The pressure signal is out of the
measurement range.
• Check the patient and treat, if necessary.
• Access the Pressure Labels menu and
assign the correct label.
• Check the equipment and replace it, if
necessary.
<xx> Please
Check Zero
The IBP zero value stored in the
monitor may not correspond to
the peripheral device.
Zero the transducer.
<xx> Static
Pressure
Static pressure was detected on a
pulsatile signal, owing to one of
the following conditions:
• A physiological condition, for
example, asystole.
• The transducer was turned off to
the patient.
• A catheter tip lodged against a
vessel wall.
• A clot on the catheter tip.
• Check the patient and treat,if necessary.
• Open the system to the patient by turning the
stopcock.
• Follow hospital procedures for dislodging
catheters.
• Follow hospital procedures for clotted
catheters.
<xx>
Unplugged
The pressure transducer for the
specified parameter is either
unplugged or defective.
• Active pressure: reconnect or replace the
cable.
• Inactive pressure: turn off alarms.
<xx> Zero
Required
The pressure transducer for the
specified parameter requires
zeroing.
Zero the transducer.
<xx> Zero
Accepted
The zeroing of the transducer was
successful
No action is required.
<xx> Did Not
Zero
The zeroing of the transducer
failed because of one of the
following conditions:
• Excessive signal noise
• A non-static waveform
• Keep all tubing motionless, then rezero.
• Change the transducer.
• Check the stopcock, then rezero.
<xx> Did Not
Zero - Offset
Error
The zeroing of the transducer
failed because static pressure
was too high or too low.
• Rezero the transducer.
• Loosen and retighten the transducer dome,
then rezero the transducer.
• Replace the transducer.
NOTE
: <xx> represents the IBP parameter label associated with the displayed message.
Содержание Infinity Delta
Страница 22: ...22 DELTA DELTA XL KAPPA VF8 INFINITY DELTA SERIES USER S GUIDE This page intentionally left blank ...
Страница 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Страница 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 102: ...4 8 DELTA DELTA XL KAPPA VF8 4 ADMISSION TRANSFER AND DISCHARGE This page intentionally left blank ...
Страница 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Страница 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Страница 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Страница 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Страница 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Страница 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Страница 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Страница 234: ...13 16 DELTA DELTA XL KAPPA VF8 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Страница 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Страница 330: ...18 16 DELTA DELTA XL KAPPA VF8 18 TRANSCUTANEOUS BLOOD GAS MONITORING This page intentionally left blank ...
Страница 358: ...20 10 DELTA DELTA XL KAPPA VF8 20 MICROSTREAM ETCO2 MONITORING This page left blank intentionally ...
Страница 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Страница 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Страница 468: ...26 30 DELTA DELTA XL KAPPA VF8 26 PULSE CONTOUR CARDIAC OUTPUT PICCO MONITORING This page intentionally left blank ...
Страница 469: ...27 Body Temperature Overview 27 2 Temperature Display 27 4 Temperature Setup 27 6 Status Messages 27 6 ...
Страница 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Страница 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Страница 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...